2.7.6. Echo Quality
2.7.6.1. Confidence Short
Short shot confidence value displays the echo confidence of the
measurements echo from the last short shot. Echo confidence measures
echo reliability. Threshold Short defines the minimum criterion for short
shot echo confidence.
Open the menu View – Profile.
2.7.6.2. Confidence Long
Long shot confidence value displays the echo confidence of the
measurements echo from the last long shot. Echo confidence measures
echo reliability. Threshold Long defines the minimum criterion for long
shot echo confidence.
Open the menu View – Profile.
2.7.6.3. Echo Strength
Displays the amplitude (in dB above 1 µV rms) of the echo selected as the
measurement echo.
Open the menu View – Profile.
Page 92
SITRANS Probe LU (PROFIBUS PA) – INSTRUCTION MANUAL
Values (view only)
Related
Parameters
Values (view only)
Related
Parameters
Values (view only)
0 to 99
----
Threshold Short
0 to 99
----
Threshold Long
–20 to 99
Shot not used
Shot not used