Risetime - HP 54501A Programming Reference Manual

Digitizing oscilloscope
Hide thumbs Also See for 54501A:
Table of Contents

Advertisement

RISetime
RISetime
command/query
The
:MEASURE:RISETIME
command places the instrument in the
continuous
measurement mode
and starts
a
Risetime measurement.
The
RISETIME
query
measures and outputs the rise
time of the
first
displayed rising (positive-going)
edge.
For
best measurement accuracy
set
the sweep speed
as
fast
as
possible while leaving the leading edge of
the
waveform
on the
display.
The
rise
time
is
determined by
measuring
the time at the lower threshold
of the rising edge then
the time
at
the
upper threshold of
the
rising edge and calculating the
rise
time with
the
formula:
rise
time
=
(time at upper threshold point
-
time
at lower threshold
point)
If
the horizontal scaling
is
questionable when
this
measurement
is
made
an "error
11" is
placed in the error queue.
:MEASure:RISetime
OUTPUT 707;":MEAS:RIS"
:MEASure:RISetime?
[:MEASure:RISetime]
<valuexNL>
Where:
<
value
>
::
=
rise
time
in
seconds (exponential
-
NR3 format)
Command
Syntax:
Example:
Query Syntax:
Returned Format:
Example:
DIM Rs$[50]
OUTPUT
707;":MEAS:RIS?"
ENTER
707;
Rs$
PRINT Rs$
HP 54501A
Programming
Measure Subsystem
14-37

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents