Problem
Test program: Test object
can now be removed
Program cancelled
Program continued
Peak load cut-out
All settings reset
All program settings reset

Door

Problem
The door is open a fraction
and cannot be closed
using the  button.
Possible cause and solution
This is not a fault.
A test program is running to check performance. At certain
points in the program, the sequence is interrupted so that
samples can be taken.
Take a sample.
or
Wait. The program will continue automatically in approx.
30 seconds.
or
Continue the program without delay by pressing the
Start/Stop button.
This is not a fault.
A program which was running was canceled by the user.
 The wash cabinet interior can be very hot.
When the door is opened, hot steam and process
chemicals can escape. Protective measures for personal
safety must be observed.
This is not a fault.
The process of canceling a program was not completed.
The program which was running continued without
interruption.
This is not a fault!
Individual components of the machine are paused while
there is a peak load signal from your energy management
system.
This is not a fault.
A user has restored factory default settings.
Confirm the message with OK.
This is not a fault!
A user has restored the factory default setting for the
program.
Confirm the message with OK.
Possible cause and solution
This is not a fault.
The Comfort door lock has opened the door slightly at the
end of the program.
Open the door. The door can now be closed completely
again using the  button.