MOBILETT
SPR8-230.621.01.03.02
Functional and Safety Checks
To ensure that the system is ready for operation and all functions relevant to
safety are working correctly, you must perform regular function and safety tests.

Daily checks

Please perform the following safety checks daily:
◆ Check the power supply plug. Do not use the MOBILETT if the power supply
plug is damaged.
◆ Check the power cable. Do not use the MOBILETT if the power cable is
damaged.
◆ Visual check of the collimator functions. Switch on the collimator lamp and
check that the light field can be adjusted with the knobs on the collimator.
◆ Check the exposure indicator lamp during exposure.
◆ Check that the audible signal indicates the end of the exposure.
◆ Check the measuring tape regarding function.
◆ Check that the tube assembly and the arm system remain in the desired posi-
tion, but are still easy to move.
◆ Check that the collimator rotates around the beam axis.
◆ For MOBILETT XP Hybrid only: Check charging status of the batteries. If low -
charge to full.
Functional and Safety Checks
71