Adjusting the radiation field

◆ Switch on the collimator light.
◆ Set the appropriate film focus distance, using the built-in measuring tape.
◆ Turn the collimator until the light field is parallel with the sides of the cassette.
◆ Place the detachable protractor on the cassette.
◆ Read the inclination angle and adjust it.
◆ Collimate the radiation field for the desired image size.
Operator Manual