Monthly checks

Please perform the following safety checks monthly:

Labels

◆ Check the legibility of the labels according to section Labels
(→ Page 99)
◆ If any labels must be replaced, please notify Siemens customer service.

DAP Measuring System (Option)

The test displays a value, default is 100.0 µGym
2 , the system should be calibrated.
80 - 120 µGym
◆ Press the Test button if you want to perform the system test.
The calibration of the chamber must be checked in accordance to national regu-
lations or laws.
2 . If the test value is not between