Printing Events

9-8
Printing Events
The HeartStart XLT can be configured to print automatically when certain
events occur. The table below lists these events and the length of the strip
printed, depending on whether the printer is configured to print real-time or
with a 6-second delay.
Table 9-4 Printing Events
Event
Defibrillator charges
Shock Delivered
Shock Failed
Defibrillator disarmed
Alarm Violation
Mark
pressed
Event
Printing is configured independently for each of these events. You can stop
the printing before it has printed the entire strip by pressing
To print additional events that you observe in the course of caring for your
Print
.
patient, press
An ECG strip will print continuously until you press
Strip
Print
a second time to stop printing. If the printer is configured to have a 6-
Strip
second delay, the printout contains an additional 6 seconds of the ECG that
occurred just prior to pressing
Real-Time
Strip Length
continuous
6 seconds just prior to charging, plus
continuous printing throughout the
charge duration.
12 seconds
6 seconds just prior to shock, plus 12
seconds after shock.
6 seconds
6 seconds just prior to the message
the message.
6 seconds
6 seconds just prior to shock, plus 6
seconds after shock.
6 seconds
6 seconds just prior to alarm viola-
tion, plus 6 seconds after alarm vio-
lation.
6 seconds
6 seconds just prior to marking, plus
6 seconds after button pressed.
Print
.
Strip
Delayed
Strip Length
, plus 6 seconds after
Print
.
Strip