Troubleshooting Tips
Table 12-3 Troubleshooting Tips (Continued)
Situation
There is a dashed (----)
line on the display
instead of an ECG.
The HeartStart XLT
does not appear to be
functioning properly.
The displayed time is
incorrect.
The printed date is
incorrect.
12-8
Cause
ECG data is not being acquired.
The battery is low.
There is a system failure.
The time was not correctly set in the con-
figuration.
The date was not correctly set in the con-
figuration.
Possible Solution
Check the patient cable is connected.
Check the pads or electrodes are prop-
erly applied.
Check that the desired lead is selected.
Change the battery.
Take the device out of use and call for
service.
Set the time in the General Settings menu
of the Configuration Mode.
Set the time in the General Settings menu
of the Configuration Mode.
Troubleshooting