Troubleshooting Tips - Philips HEARTSTART XLT M3500B Instructions For Use Manual

Defibrillator/monitor
Table of Contents

Advertisement

Table 12-3 Troubleshooting Tips
Situation
The HeartStart XLT
does not turn on.
The display is blank
except for
" SYSTEM FAILURE
CYCLE POWER ".
The HeartStart XLT
appears to be on but the
display is blank.
M3500B HeartStart XLT Defibrillator/Monitor

Troubleshooting Tips

Table 12-3 lists some situations that you may encounter, causes, and actions to
take.
Cause
There is no power.
OR
A corrupt Data Card may prevent the unit
from powering on.
OR
Hardware failure.
An internal error occurred.
The contrast needs adjusting.
The display has overheated.
Troubleshooting Tips
Possible Solution
Insert a fully charged battery.
Attach AC Power cord.
Remove the Data Card
Insert a new Data Card, if available
Turn on the device
Remove this unit from clinical use and
call for service.
If this occurs during actual use:
Substitute another defib, if possible.
Remove this unit from clinical use and
call for service.
If no other defib is available, turn
power off, then on. If unit turns on
normally, use for this one incident.
After this incident, remove from clini-
cal use and call for service.
If unit does not turn on normally, it
cannot be used.
If this occurs during routine testing:
Remove this unit from clinical use and
call for service.
Adjust the contrast.
Avoid exposing the display to direct
sunlight on a hot day.
12
12-7