Pacing

8-4
Pacing
To perform pacing:
1. Press
Pacer
. The green LED next to
logue box appears.
Figure 8-1 Pacing Display
LEAD
HR ALARM
SELECT
78
Lead II
Pacer Stop
Demand Mode
150 J
70ppm 30mA
ENERGY
SELECT
The message
are not being delivered. The pacer turns on in the mode last used.
2. Verify that the dot markers appears near the middle of the QRS complexes
of the ECG.
If the dot markers do not appear, or are in the wrong location, adjust the
ECG size or select another lead. (See "Monitoring the ECG" in Chapter
4.)
The ECG will only appear if the 3- or 5-lead monitoring cable and elec-
trodes are used.
3. Press
to change to fixed mode, if R-wave detection is unreliable.
Mode
The pacing dialogue box displays the current mode. To switch back to
demand mode, press
SpO2
Shocks:0
00:00:8
CHARGE
indicates that the pacing function is on but paced pulses
again.
Mode
Pacer
lights up and the dia-
Pacing