Connecting The H Solution Container - Miele PG 8527 Operating Instructions Manual

Washer-disinfector
Table of Contents

Advertisement

Special processes - OXIVARIO and ORTHOVARIO
Connecting the H
container
The connection hose for the H
solution container is identified by a
black label. The connection hose is
supplied without a safety cap as the
container for hydrogen peroxide will
vary depending on supplier.
^ Connect the supplier's safety cap to
the connection hose (black).
^ Then connect this to the H
solution container.
^ Then start the
DOS2-FILL
programme.
Unlike with other chemical agents the
H
O
solution container must be fully
2
2
emptied before it is exchanged for a
new one.
^ Do not replace the H
container with a new one until the
DOS2 container
displayed. After connecting the new
container, you must run the
DOS2-FILL
programme.
^ If the
Check dispensing system 2
message appears you should check
the container as well as the
dispensing system. The programme
will stop automatically.
O
solution
2
2
O
2
2
O
2
2
O
solution
2
2
message is
Fill
79