Special processes - OXIVARIO and ORTHOVARIO
The ORTHOVARIO procedure is
suitable for reprocessing aluminium
instruments which are sensitive to
alkalines. It is particularly suitable for
orthopaedic instruments including
motor systems.
Because of its oxidising effect it
must not be used for titanium
implants, and in particular colour
coded ones.
,
Due to the high efficacy of the
Oxivario and Orthovario processes,
it is important to lubricate joints and
contact parts on instruments
pre-sterilisation. Please refer to the
instrument manufacturer for
guidance.
78
Warning and Safety
instructions
,
These warning and safety
instructions are in addition to those
given at the beginning of this
booklet.
~
H
O
solution must only be used in
2
2
the special containers and adapters
provided by the Ecolab or Dr. Weigert
companies.
~
Please observe safety instructions
(safety data sheets) provided by the
chemical agent manufacturer.
~
Be especially careful when handling
H
O
solution. It is an irritant chemical.
2
2
Observe all current safety instructions
pertaining to its use.
Wear protective gloves and goggles.
~
Empty containers must be disposed
of in accordance with the
manufacturer's instructions.
~
H
O
solution must not be mixed
2
2
with other chemicals. This could cause
a serious chemical reaction, e. g. the
release of harmful vapours or gases.
~
Only use tenside cleaning agents
supplied by Ecolab or Dr. Weigert.