Maintenance; Routine Checks - Miele G 7883 Operating And Installation Instructions

Laboratory glassware washer
Table of Contents

Advertisement

Maintenance

Periodic maintenance for the lab
washer is recommended annually.

Routine checks

The routine checks must be done daily
by the user before using the machine.
For the routine checks a checklist is
supplied with the machine.
Check the following points:
– the filters in the base of the wash
cabinet must be empty,
– the spray arms in the machine and at
the baskets should be clear,
– the wash cabinet and the door
gasket should be clear,
– the dispensing systems, and
– the baskets and inserts.
48
When using powder detergent the
temperature at the time of dosing must
be checked every 2 weeks. The dosing
temperature is found under Validation
in the Validation Protocol.
The temperature review must be
completed during a running program,
when the container lid can be heard
opening. The temperature in the display
can be compared with that shown in
the validation protocol.
,
If the temperature is found to be
more that 2° plus or minus the
required temperature, please
contact Miele.