Sanyo VR-L4110WSEC - Commercial Solutions Lab Manual page 27

Booklet
Table of Contents

Advertisement

Integrated LCD control with graphical display. The Twin Guard Series
microprocessor controller with LCD information center to simplify all freezer functions. Uniform ultra-low temperature
is achieved through a combination of performance systems supervised by the controller complete with alarm,
programming and diagnostic protocols.
The Safest Ultra-Low Freezer
for Long-Term Storage of Ultra-
Critical Biologicals
The SANYO Twin Guard Series
the industry demand for safe, long-
term storage for the most high-valued
materials. Two independent refrigeration
systems, combined with optional liquid
nitrogen or liquid CO
back-up systems,
2
offer a circle of protection unmatched in
the marketplace. Developed for use with
conventional inventory racks and boxes,
the Twin Guard Series
is ideal for stor-
age of sensitive stem cells, embryos,
cell lines, and other rare specimens.
Independent systems efficient ultra-low cooling
is achieved through two independent evaporator
circuits surrounding the interior chamber.
MDF-U500VXC is managed by an integrated
Twin Guard Series
Refrigeration System
The Twin Guard Series
er avoids conventional cascade refrigera-
satisfies
tion technology by using two completely
independent one-compressor, autocas-
cade cooling systems, each capable of
maintaining ultra-low temperatures.
Each refrigeration circuit includes a
closed-loop cold-wall evaporator con-
figured in parallel to the other.
Independent evaporators and cooling
fans assure back up status at all times,
eliminating system failure due to
sub-component failure in conventional
cascade systems configured of mutu-
ally dependent high- and low-stage
systems.
Failure Mode Comparison
Event
Fan Failure
High Stage
Compressor
Failure
Low Stage
Compressor
Failure
-86 Dual°Cool
Ultra-Low Freez-
Conventional Freezer
Conventional Freezer
No Backup
With Backup
Freezer Fails
Freezer fails,
CO
or LN
2
system offers
short-term
Freezer Fails
protection until
contents can
be removed and
repairs initiated.
Freezer Fails
In the unlikely event of a compressor
failure in one system, the remaining
system will automatically maintain a
minimum of -65°C for an indefinite
period.
In the event of a facility power failure
with optional CO
or LN
backup sys-
2
2
tem installed, the freezer will maintain
-65°C storage temperature for up to
eight hours (CO
backup system) and
2
-65°C storage temperature for up to
15 hours (LN
backup system).
2
A unique EcoMode
deploys both sys-
tems in overlapping cycles to maintain
-86°C and to reduce energy consump-
tion by as much as 15%.
Evaporator coils embedded in the
patented, high-tech, SANYO V.I.P .
vacuum-insulated thin-wall cabinet are
strategically oriented to deliver the
best temperature uniformity at all shelf
levels, top-to-bottom and front-to-back.
New SANYO-designed Cool Safe
refrigeration compressors feature inno-
vative refrigerant feedback processes
to reduce compressor temperature,
thereby extending compressor life and
minimizing heat output.
SANYO
Twin Guard
If one fan fails, the second fan auto-
matically maintains energy exchange.
backup
2
No high or low stage used. Two refrig-
eration systems, each with a single
compressor, operate independently in
overlapping cycles during normal opera-
tion. If one compressor fails internal
temperature is maintained indefinitely
at -65°C. Optional LN
or CO
2
system offers additional protection.
Series
backup
2