Sanyo VR-L4110WSEC - Commercial Solutions Lab Manual page 16

Booklet
Table of Contents

Advertisement

CO
/O
Laboratory Incubators
2
2
SANYO models MCO-5M/18M/36M series CO
environment. MCO Series CO
and CO
2
applications in accordance with the FDA Safe Medical Devices Act of 1990 and the Medical Device Amendments of 1992.*
CO
/O
Laboratory Incubators
2
2
Volume
Exterior Dimensions
Model Number
(cu.ft.)
1.7
18.9" x 21.6" x 22.6"
MCO-5M
49 L
480 x 548 x 575 mm
6.0
24.4" x 28.0" x 35.4"
MCO-18M
170 L
620 x 710 x 900 mm
12.0
24.4" x 28.0" x 70.8"
MCO-36M
340 L
620 x 710 x 1800 mm
For below ambient or enriched (above ambient) oxygen levels in addition to CO
16
Guide to Laboratory Equipment www.sanyobiomedical.com
/O
incubators employ multiple sensor technologies to achieve in vitro simulation of the in vivo
2
2
/O
incubators have received U.S. Food and Drug Administration 510(k) clearance for in vitro fertilization
2
2
(w x f-b x h)
Contamination Control
inCu saFe copper-enriched
stainless steel interior,
optional SafeCell UV with
ultraviolet light.
control.
2
Infrared CO
Sensor
2
Precise CO
control, fast response to
2
door openings. The SANYO CO
sor monitors and controls CO
over a range of 0% to 20%, with control
of ±0.15%.
Inner Doors
Multiple chamber inner doors minimize
loss of balanced interior atmosphere dur-
ing routine door openings (available on
selected models).
Professional Cell Culture
Multi-Gas Incubator:
Continuous contamination
control with inCu saFe
interior and SafeCell
technologies.
P .I.D. controls for fast
recovery of temperature,
CO
and O
2
Fast humidity level recovery
by N
2
CO
Control
2
thermal conductivity
infrared with
P.I.D./R recovery
Zirconia O
Non-depleting design for precise O
IR sen-
control with fast response to door
2
level(s)
openings. The maintenance-free zirconia
2
solid-state electrolyte sensor has a high
degree of precision, a long service life
and does not require fine adjustment.
Through accurate determination of the
chamber O
injects either nitrogen gas or oxygen
as required.
P .I.D./R
Proportional, integral and derivative infra-
red CO
prevents overshoot.
MCO-18M
*
UV
levels.
2
gas bubbler.
Voltage, Power
O
Control
Connection
2
Zirconia sensor with
P.I.D./R recovery
NEMA 5-15
Control
2
level the microprocessor
2
Control Sophistication
control accelerates recovery and
2
Reference Number: K013703
Regulation Number: 21 CFR 884-6120, Assisted
Reproduction Accessories, Regulatory Class II,
Product Code 85MOG, October 30, 2001.
MCO-5M
(shown stacked)
MCO-5M
115V