User Manual
Implant Center 2

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Implant Center 2 and is the answer not in the manual?

Questions and answers

Summary of Contents for Acteon Implant Center 2

  • Page 1 User Manual Implant Center 2...
  • Page 2 This document is an English translation of the original French version. Reference J27171FR version V1 and drawing number NO20FR010A...
  • Page 3: Table Of Contents

    Contents 1 Documentation 1.1 Associated documentation 1.2 Electronic documentation 2 Required information 2.1 Intended use 2.2 Indication for use 2.3 Operating principle 2.4 Using accessories not supplied by the manufacturer 2.5 Connecting and disconnecting accessories during use 2.6 Repairing or modifying the medical device 2.7 Warranty 2.8 Latest document update 2.9 Date of first CE marking...
  • Page 4 7.2.2.1 Piezotome mode 7.2.2.2 Newtron mode 7.2.2.3 I-surge mode 7.3 Accessories 7.3.1 Handpiece 7.3.2 Handpiece support 7.3.3 Handpiece cord 7.3.4 Irrigation lines 7.3.5 Motor 7.4 Adjustments 7.4.1 I-Surge mode factory settings 7.4.2 Ultrasound power 7.4.2.1 Piezotome mode factory settings 7.4.2.2 Newtron mode factory settings 7.4.3 Irrigation 7.4.4 Adjusting the medical device volume 7.4.5 Adjusting the screen brightness...
  • Page 5 12.3 Symbols 12.4 Quick Start and Quick Clean symbols 12.5 Manufacturer identification 12.6 Branch addresses 12.7 Disposal and recycling 13 Index...
  • Page 7: Documentation

    All paper or electronic documentation relating to your medical device must be kept for the device’s entire service life. User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 5/62...
  • Page 8 Keep the original documentation for your medical device and its accessories for future reference. When loaning out or selling the medical device, the documentation must be provided with it. Page 6/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 9: Required Information

    2 Required information 2.1 Intended use The intended use of the SATELEC Implant Center 2 is to provide utilities and to act as a base for dental instruments and accessories used by professionals qualified in dentistry. 2.2 Indication for use The Implant Center 2 is a command unit used in combination with the following items.
  • Page 10: Warranty

    The user must not remove any of the screws shown in this view or risk invalidating the medical device's warranty. 2.8 Latest document update 04/2022 2.9 Date of first CE marking 2010 Page 8/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 11: Unpacking The Medical Device

    Plastic clips for the irrigation line An attachment kit. Tips and wrenches depending on selected options. A power cord Two brackets to hold the irrigation solutions User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 9/62...
  • Page 12 Page 10/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 13: Connect The Medical Device

    1. Set the medical device's mains switch to "O" OFF position. 2. Connect the mains cord to the control unit's mains connector. 3. Connect the power lead to the mains socket. User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 11/62...
  • Page 14 Page 12/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 15: Installing The Medical Device

    Align the mark on the cord connector with the mark on the plug. Insert the connector until you hear a click. The Implant Center 2 enables the connection of either an intraoral dental surgery handpiece or a dental handpiece. User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 13/62...
  • Page 16: Connecting The Motor

    9. On the short infusion line side, pierce the irrigation solution bag with the perforator. 10. When the medical device is switched ON, open the perforator cap and purge the irrigation system. Page 14/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 17: Attaching A Tip Or A File

    II on the base. They are incompatible with first-generation Implant Center and Piezotome tips. Conversely, older-generation tips are incompatible with Piezotome Cube, Implant Center Cube, Implant Center 2, and Piezotome Solo LED. User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 15/62...
  • Page 18 Page 16/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 19: Dispensing A Treatment

    Do the same for setting up the irrigation line for the micromotor, steps 1 to 13. Your medical device is now ready to use. User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 17/62...
  • Page 20: I-Surge Mode

    6.3 I-surge mode 6.3.1 Choosing a contra-angle Page 18/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 21 User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 19/62...
  • Page 22: Select The Speed Of Rotation

    6.3.2 Select the speed of rotation Page 20/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 23 User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 21/62...
  • Page 24: Choosing The Torque

    6.3.3 Choosing the torque Page 22/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 25 User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 23/62...
  • Page 26: Piezotome Mode

    6.4 Piezotome mode Page 24/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 27: Newtron Mode

    6.5 Newtron mode User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 25/62...
  • Page 28: Switching Off The Medical Device

    When not in use, or in storage or before a long absence, disconnect the medical device from the mains power supply. Page 26/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 29: Medical Device Description

    The control unit incorporates electronic control elements to drive a brushless sensorless motor. 7.1.1 Front view of the medical device 7.1.2 View of the right-hand side of the medical device User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 27/62...
  • Page 30: Rear View Of The Medical Device

    It is tactile such that practitioners can make adjustments by pressing the active areas. Always make adjustments with your finger. Never use stylets or instruments, as these may damage the screen. Page 28/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 31 Warning. The type of warning is specified by a pictogram at the centre of the triangle. Save any changes made to programmes User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 29/62...
  • Page 32 Smooth footswitch Pictograms for the Newtron Meaning Power mode Soft programme Medium programme Medium High programme High Boost programme Very high Page 30/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 33 Select programme P1 - Marking of the implant site Select programme P2 - Pilot drilling Select programme P3 - Boring/threading Select programme P4 - Screwing Clockwise rotation Anti-clockwise rotation User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 31/62...
  • Page 34: Piezotome Mode

    7.2.1.1 Piezotome mode Page 32/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 35: Newtron Mode

    7.2.1.2 Newtron mode User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 33/62...
  • Page 36: I-Surge Mode

    D2, D3 and D4. In I-Surge mode, press to cycle through programmes P1, P2, P3 and P4. Press to enable or disable irrigation. Page 34/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 37: Piezotome Mode

    Switch from motor to ultrasound and vice versa. Activate the purge. I-Surge mode. Activate motor reciprocation. 7.2.2.1 Piezotome mode 7.2.2.2 Newtron mode User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 35/62...
  • Page 38: I-Surge Mode

    The autoclavable irrigation lines can be reused after cleaning, disinfection and sterilisation. After use, sterile irrigation lines should be discarded in a biomedical waste container. Page 36/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 39: Motor

    60 Hz Programme Fine setting level Power value 64 % 68 % 72 % 74 % 78 % Frequency modulation 60 Hz User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 37/62...
  • Page 40: Newtron Mode Factory Settings

    1. Switch on the medical device. 2. From the home screen, press to open the menu. 3. Identify the pictogram indicating brightness Page 38/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 41: Adjusting The Lighting Timer

    Any customised contra-angle and any changes made to programmes will be deleted from the medical device's memory. Or press to exit the settings screen without saving the changes. User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 39/62...
  • Page 42 Page 40/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 43: Disinfection And Sterilising

    The autoclavable irrigation lines must be cleaned, disinfected and sterilised after each use. To prepare for cleaning, remove the various parts of the Implant Center 2 as shown here. User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 41/62...
  • Page 44: Cleaning, Disinfecting And Sterilising Accessories

    8.2 Cleaning, disinfecting and sterilising accessories Refer to the cleaning, disinfection and sterilisation instructions for accessories listed in the chapter Associated documentation page 5. Page 42/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 45: Monitoring And Routine Maintenance

    9.2.2 Missing handpiece Symptom: the medical device beeps when the user presses the footswitch and displays the following pictogram: User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 43/62...
  • Page 46: I-Surge Led Micromotor Missing

    4. Insert the tip of a flathead screwdriver into the notch on top of the fuse holder to release it. 5. Remove the used fuses. Page 44/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 47 7. Place the fuse holder in its recess by pushing it until you hear a click that confirms it is in the correct position. 8. Connect the mains cord to the connector. 9. Connect the mains cord to the electrical network. User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 45/62...
  • Page 48 Page 46/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 49: Identifying Incorrect Operation

    Unsuitable contra-angle Change the contra-angle 10.4 No spray Symptom: There is no water spray at the tip. User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 47/62...
  • Page 50: The Power Is Not As Expected

    Satelec's Customer Service. Pipe ruptured in the irrigation line cassette Replace the irrigation line Irrigation cassette not working Replace the irrigation line Page 48/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 51: Technical Specifications Of The Medical Device

    Length including nosepiece 90.6 mm Length excluding nosepiece 69 mm Diameter of the nosepiece 10 mm max. Diameter 23.2 mm Weight 115 g User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 49/62...
  • Page 52: Irrigation

    Immersion The console must not be immersed. Immersion The handpiece must not be immersed. Page 50/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 53: Main Performance Characteristics

    Ultrasonic vibrations of the tip or file fitted to the end of the conventional dental ultrasonic handpiece. Vibration frequency ≥ 28 kHz. Modulation frequency User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 51/62...
  • Page 54 Page 52/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 55: Regulations And Standards

    Refer to the supporting documentation Consult the User Manual The accompanying documentation is available in electronic format Pressure limit Temperature limit Humidity limit User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 53/62...
  • Page 56 Type BF part in contact Class 1 Alternating current CE marking CE marking Year of manufacture YYYY Manufacturer Do not dispose of as household waste Page 54/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 57 Use a soft brush for cleaning Use a lint-free cloth for cleaning Use an ultrasonic tank for cleaning. Use a swab for cleaning User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 55/62...
  • Page 58 Australia RUSSIA Tel. +612 9669 477 307 Mob. +7 926 233 1695 Fax. +612/96692204 Tel. +7 495 150 1323 info.au@acteongroup.com info.ru@acteongroup.com Page 56/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 59 Mount Laurel, NJ 08054 - USA Mobile: +57 312 377 8209 Tel. +1 856 222 9988 info.latam@acteongroup.com Fax. +1 856 222 4726 info.us@acteongroup.com User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 57/62...
  • Page 60 ACTEON MIDDLE EAST 247 Wasfi Al Tal str. 401 AMMAN - JORDAN Tel. +962 6 553 4401 Fax. +962 6 553 7833 info.me@acteongroup.com Page 58/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 61 When your medical device has reached the end of its service life, contact your nearest dental equipment dealer, or the Acteon head office or one of the company branches to find out how to proceed. The relevant contact details are given in the chapter Branch addresses page 56.
  • Page 62 38, 47 gas-filled atmosphere 50 Sterile water 37 sterilised 17 switch 28, 47 syndesmotomy 7 handpiece 5 Handpiece support 36 Page 60/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
  • Page 63 50 tip 5, 7, 47-48 ultrasonic vibrations 7 update 8 User Manual 5 Vibration frequency 51 water leakage 48 User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 61/62...
  • Page 64 User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A SATELEC S.A.S. |A Company of ACTEON Group 17 av. Gustave Eiffel | BP30216 | 33708 MERIGNAC cedex | FRANCE Tel. +33(0)556340607 | Fax. +33(0)556349292...