Contents 1 Documentation 1.1 Associated documentation 1.2 Electronic documentation 2 Required information 2.1 Intended use 2.2 Indication for use 2.3 Operating principle 2.4 Using accessories not supplied by the manufacturer 2.5 Connecting and disconnecting accessories during use 2.6 Repairing or modifying the medical device 2.7 Warranty 2.8 Latest document update 2.9 Date of first CE marking...
Page 4
7.2.2.1 Piezotome mode 7.2.2.2 Newtron mode 7.2.2.3 I-surge mode 7.3 Accessories 7.3.1 Handpiece 7.3.2 Handpiece support 7.3.3 Handpiece cord 7.3.4 Irrigation lines 7.3.5 Motor 7.4 Adjustments 7.4.1 I-Surge mode factory settings 7.4.2 Ultrasound power 7.4.2.1 Piezotome mode factory settings 7.4.2.2 Newtron mode factory settings 7.4.3 Irrigation 7.4.4 Adjusting the medical device volume 7.4.5 Adjusting the screen brightness...
All paper or electronic documentation relating to your medical device must be kept for the device’s entire service life. User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 5/62...
Page 8
Keep the original documentation for your medical device and its accessories for future reference. When loaning out or selling the medical device, the documentation must be provided with it. Page 6/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
2 Required information 2.1 Intended use The intended use of the SATELEC Implant Center 2 is to provide utilities and to act as a base for dental instruments and accessories used by professionals qualified in dentistry. 2.2 Indication for use The Implant Center 2 is a command unit used in combination with the following items.
The user must not remove any of the screws shown in this view or risk invalidating the medical device's warranty. 2.8 Latest document update 04/2022 2.9 Date of first CE marking 2010 Page 8/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
Plastic clips for the irrigation line An attachment kit. Tips and wrenches depending on selected options. A power cord Two brackets to hold the irrigation solutions User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 9/62...
1. Set the medical device's mains switch to "O" OFF position. 2. Connect the mains cord to the control unit's mains connector. 3. Connect the power lead to the mains socket. User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 11/62...
Align the mark on the cord connector with the mark on the plug. Insert the connector until you hear a click. The Implant Center 2 enables the connection of either an intraoral dental surgery handpiece or a dental handpiece. User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 13/62...
9. On the short infusion line side, pierce the irrigation solution bag with the perforator. 10. When the medical device is switched ON, open the perforator cap and purge the irrigation system. Page 14/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
II on the base. They are incompatible with first-generation Implant Center and Piezotome tips. Conversely, older-generation tips are incompatible with Piezotome Cube, Implant Center Cube, Implant Center 2, and Piezotome Solo LED. User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 15/62...
Do the same for setting up the irrigation line for the micromotor, steps 1 to 13. Your medical device is now ready to use. User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 17/62...
When not in use, or in storage or before a long absence, disconnect the medical device from the mains power supply. Page 26/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
The control unit incorporates electronic control elements to drive a brushless sensorless motor. 7.1.1 Front view of the medical device 7.1.2 View of the right-hand side of the medical device User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 27/62...
It is tactile such that practitioners can make adjustments by pressing the active areas. Always make adjustments with your finger. Never use stylets or instruments, as these may damage the screen. Page 28/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
Page 31
Warning. The type of warning is specified by a pictogram at the centre of the triangle. Save any changes made to programmes User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 29/62...
Page 32
Smooth footswitch Pictograms for the Newtron Meaning Power mode Soft programme Medium programme Medium High programme High Boost programme Very high Page 30/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
Page 33
Select programme P1 - Marking of the implant site Select programme P2 - Pilot drilling Select programme P3 - Boring/threading Select programme P4 - Screwing Clockwise rotation Anti-clockwise rotation User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 31/62...
D2, D3 and D4. In I-Surge mode, press to cycle through programmes P1, P2, P3 and P4. Press to enable or disable irrigation. Page 34/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
The autoclavable irrigation lines can be reused after cleaning, disinfection and sterilisation. After use, sterile irrigation lines should be discarded in a biomedical waste container. Page 36/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
1. Switch on the medical device. 2. From the home screen, press to open the menu. 3. Identify the pictogram indicating brightness Page 38/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
Any customised contra-angle and any changes made to programmes will be deleted from the medical device's memory. Or press to exit the settings screen without saving the changes. User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 39/62...
The autoclavable irrigation lines must be cleaned, disinfected and sterilised after each use. To prepare for cleaning, remove the various parts of the Implant Center 2 as shown here. User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 41/62...
8.2 Cleaning, disinfecting and sterilising accessories Refer to the cleaning, disinfection and sterilisation instructions for accessories listed in the chapter Associated documentation page 5. Page 42/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
9.2.2 Missing handpiece Symptom: the medical device beeps when the user presses the footswitch and displays the following pictogram: User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 43/62...
4. Insert the tip of a flathead screwdriver into the notch on top of the fuse holder to release it. 5. Remove the used fuses. Page 44/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
Page 47
7. Place the fuse holder in its recess by pushing it until you hear a click that confirms it is in the correct position. 8. Connect the mains cord to the connector. 9. Connect the mains cord to the electrical network. User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 45/62...
Unsuitable contra-angle Change the contra-angle 10.4 No spray Symptom: There is no water spray at the tip. User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 47/62...
Satelec's Customer Service. Pipe ruptured in the irrigation line cassette Replace the irrigation line Irrigation cassette not working Replace the irrigation line Page 48/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
Length including nosepiece 90.6 mm Length excluding nosepiece 69 mm Diameter of the nosepiece 10 mm max. Diameter 23.2 mm Weight 115 g User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 49/62...
Immersion The console must not be immersed. Immersion The handpiece must not be immersed. Page 50/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
Ultrasonic vibrations of the tip or file fitted to the end of the conventional dental ultrasonic handpiece. Vibration frequency ≥ 28 kHz. Modulation frequency User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 51/62...
Refer to the supporting documentation Consult the User Manual The accompanying documentation is available in electronic format Pressure limit Temperature limit Humidity limit User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 53/62...
Page 56
Type BF part in contact Class 1 Alternating current CE marking CE marking Year of manufacture YYYY Manufacturer Do not dispose of as household waste Page 54/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
Page 57
Use a soft brush for cleaning Use a lint-free cloth for cleaning Use an ultrasonic tank for cleaning. Use a swab for cleaning User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A - Page 55/62...
Page 60
ACTEON MIDDLE EAST 247 Wasfi Al Tal str. 401 AMMAN - JORDAN Tel. +962 6 553 4401 Fax. +962 6 553 7833 info.me@acteongroup.com Page 58/62 - User Manual | Implant Center 2 | J27171EN | V1 | (10) | 04/2022 | NO20EN010A...
Page 61
When your medical device has reached the end of its service life, contact your nearest dental equipment dealer, or the Acteon head office or one of the company branches to find out how to proceed. The relevant contact details are given in the chapter Branch addresses page 56.