Set Up the Cleaver
Adjusting the clamping force
The clamping force of the CT114 can be adjusted by the changing the magnet positions and
tightening the clamp lid screw.
There are 4 magnets in the clamp lid.
The magnetic force can be changed
by changing their positions.
A torque driver capable of measuring in centinewton-meters (cNm) is required
for tightening the clamp lid screw.
Change the position of the magnets according to the cleaving requirements.
Magnets
CT114_Rev2
The clamp lid screw is used for
fibers with a cladding diameter
between 400μm and 660μm.
Clamp lid screw