Erbe ESM 2 Service Manual
Hide thumbs Also See for ESM 2:

Advertisement

SERVICE MANUAL
ESM 2
V 1.0.x
HYDROSURGERY
80116-294_V24157
2022-07

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ESM 2 and is the answer not in the manual?

Questions and answers

Summary of Contents for Erbe ESM 2

  • Page 1 SERVICE MANUAL ESM 2 V 1.0.x HYDROSURGERY 80116-294_V24157 2022-07...
  • Page 3 SERVICE MANUAL ESM 2...
  • Page 4 (by photocopying, microfilming or other methods) or processed, duplicated or dissem- inated by the use of electronic systems without the written consent of Erbe Elektromedizin GmbH. The information contained in this manual may be amended or supplemented without prior notice and represents no obli- gation on the part of Erbe Elektromedizin GmbH.
  • Page 5: Table Of Contents

    Table of Contents Table of Contents Safety information ........7 Classification of the safety information.
  • Page 6 Table of Contents / 24...
  • Page 7: Safety Information

    The supply voltage must match the voltage specified on the rating plate. Connect the unit / the equipment cart to a properly installed grounded outlet. Please use the Erbe power cord only; you can find the REF Numbers in the User Manual of the unit.
  • Page 8: Electrostatically Sensitive Components

    Liability and warranty ATTENTION! Adjustments, tests, modifications, maintenance and repair work may only be performed by Erbe or persons trained by Erbe. If the work is not performed by trained persons, Erbe accepts no liability and warranty rights become void.
  • Page 9: Controls

    This chapter contains an overview of the controls of the unit(s). The rel- evant User Manual for the unit(s), knowledge of which is assumed for servicing work, provides detailed information about how to use the unit(s). Controls at the front ESM 2 Fig. 2-1 Power Switch Power LED Ready LED...
  • Page 10: Controls At The Rear

    2 • Controls Controls at the rear Fig. 2-2 Silencer Vacuum connection Power supply module with fuses Potential equalization terminal ECB socket (Erbe Communication Bus) / 24...
  • Page 11: Technical Data

    3 • Technical Data CHAPTER 3 Technical Data Power connection Rated supply voltage 100 V to 240 V (± 10 %) Rated supply frequency 50 / 60 Hz Line current 0.4 A to 0.8 A Power input in standby mode 25 watts Maximum power consumption 100 watts...
  • Page 12 3 • Technical Data Ambient conditions for operation of unit Maximum operating altitude 5000 m above sea level Note: The operating conditions may be limited by the suction accessories. Note the operating conditions for the suction accessories. Acclimatizing If the unit has been stored or transported at temperatures below +10 °C or above +40 °C, the unit will require approx.
  • Page 13: Setup And Service Settings

    4 • Setup and Service settings CHAPTER 4 Setup and Service settings The Setup and service settings for the ESM 2 are called up and modified via the ERBEJET 2 water jet surgical unit. They are described in the Service Manual of the ERBEJET 2.
  • Page 14 4 • Setup and Service settings / 24...
  • Page 15: Remedying Malfunctions

    Safety information ATTENTION! Adjustments, technical tests, modifications, maintenance and repair work may only be performed by Erbe or persons trained by Erbe. If the work is not performed by trained persons, Erbe accepts no liability and warranty rights become void.
  • Page 16 5 • Remedying malfunctions Recognizing module: S = ESM 2 Recog- Error code Description Remedy nizing module CRC error. Switch the device off and on again. If the error remains inform Erbe Service. Inform Erbe Service if this occurs fre- quently.
  • Page 17 5 • Remedying malfunctions Recognizing module: S = ESM 2 Recog- Error code Description Remedy nizing module Motor current too low. Switch the device off and on again. If the error remains inform Erbe Service. Inform Erbe Service if this occurs fre- quently.
  • Page 18: Other Faults

    5 • Remedying malfunctions Recognizing module: S = ESM 2 Recog- Error code Description Remedy nizing module A current flows at the valve even Switch the device off and on again. If though none should flow. The tran- the error remains inform Erbe Service.
  • Page 19: Maintenance And Servicing

    ATTENTION! Adjustments, tests, modifications, maintenance and repair work may only be performed by Erbe or persons trained by Erbe. If the work is not performed by trained persons, Erbe accepts no liability and warranty rights become void. It is recommended that the technical safety check also be performed by Erbe or persons trained by Erbe.
  • Page 20: Testing And Measuring Equipment

    Testing and measuring equipment IMPORTANT! The following list contains the testing and measuring equipment rec- ommended by Erbe for servicing. Where Erbe article numbers are spec- ified, only original Erbe testing and measuring equipment should be used. Erbe Art. No.
  • Page 21: Technical Safety Check - Step By Step

    IMPORTANT! The test report for the technical safety check can be requested from Erbe Technical Service Tübingen. For the address see address sheet on last page. User manual and visual inspections • Test specimen and accessories (where enclosed) undamaged exter- nally.
  • Page 22: Performance Tests

    Test setup • The test specimen is connected to the ERBEJET 2 via the ECB. • The ESM 2 and ERBEJET 2 are connected to a power supply network and switched on via one power supply cable each. • The single-pedal footswitch is connected to the ERBEJET 2.
  • Page 23: Test For The Pressure Control

    6 • Maintenance and servicing Test for the pressure control Test setup ERBEJET 2 51700-252 40cm ESM 2 55701-082 20340-004 20340-012 / 20340-013 20340-102 / 20340-103 Fig. 6-1 • The test specimen is connected to the ERBEJET 2 via the ECB.
  • Page 24 6 • Maintenance and servicing / 24...