Chapter 9 Reprocessing the Endoscope (and related reprocessing accessories)
1.
Turn the elevator control lever all the way in the opposite direction of the
"
U" direction. perform the following brushing in the detergent solution.
2.
While holding the distal end, brush the groove of the interior of the forceps
elevator with the cleaning brush (MAJ-1534) until all debris is removed (see
Figure 9.9).
Cleaning brush
(MAJ-1534)
Figure 9.9
3.
Look at the left side of the instrument channel opening from the distal end,
brush the interior of this part with the cleaning brush (MAJ-1534) until all
debris is removed (see Figure 9.10).
Cleaning brush
(MAJ-1534)
Figure 9.10
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Forceps elevator
141