6.9
OER-AW (Olympus Endoscope Reprocessor)
*1
OER-AW
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
When using Olympus OER-AW, move the elevator control
lever to make the instrument channel outlet of the distal end
opened (i.e., lower the forceps elevator), attach the dedicated
connecting tube to the distal end of the endoscope, and set it
in your OER-AW according to its instructions. When using an
AER/WD that is not recommended by Olympus, set the
forceps elevator according to the AER/WD manufacturer's
instructions.
Elevator
control lever
Figure 6.1
•
OER-AW is intended to clean and disinfect endoscopes and their
accessories.
•
OER-AW is not available in the member states of the EU.
•
When cleaning and disinfecting the endoscope in OER-AW,
use connectors/retaining rack that are compatible with the
endoscope model. The compatible connectors/retaining rack
for the endoscope model should be listed in the "List of
Compatible Endoscopes/Connecting Tubes" and the
instruction manual for OER-AW or Table 6.11.
Connectors
For
For
air/water
instrument
channel
channel
MAJ-1517
Table 6.11
Chapter 6 Compatible Reprocessing Methods
For
For
forceps
suction
elevator
channel
channel
MAJ-1971
Forceps
elevator
Instrument
channel outlet
Retaining rack
MAJ-840
99