8.7. Specifying Qualification Parameters

Specify the conditions that control the qualification study on the Qualification Parameters
screen. From this screen, specify:
Qualification start / stop conditions - used to start and stop the qualification cycle
manually or automatically.
Data storage options - the rates at which data is written to the memory of the Logger
during a qualification run. There are three options, Sampling Rate, RF Transmission
Rate, and RF Threshold.
Start / Stop Conditions
Loggers may be programmed to start data collection Manually when the Qualification is
started or based on the Date / Time of Day. Loggers may be programmed to stop collection
Manually or based on Cycle time. If the Users selects start Qual on time, they can enter a date
and time up to three days from the current time. When the Setup is saved, the software
prompts the User if they would like to edit the time before starting the Qual. If yes is selected,
then the User can edit the time just prior to programming the Loggers.
Qualification Start / Stop Conditions and Related Delays:
Time of Day cannot be programmed for more than 72 hours after Loggers are
programmed. After 72 hours, the Loggers are set to shelf mode if a study is not started.
Note: When programming the start on time of the day, program one sample before the
required first sample timestamp. E.g., If the first sample is required at 14:30:00 and the
sample rate is 30 seconds, program the study start for 14:29:30.
Cycle Time does not have any duration limit, cycle time duration is dependent on
sampling rate and the number of samples. The maximum number of collected samples
is 100,000 per sensor.
ValProbe RT User Manual
Figure 104: Qualification Parameters
Options Qual Start
1
Manual
2
Manual
3
Time of Day
4
Time of Day
Chapter 8: Defining Study Setups
Qual Stop
Manual
Cycle Time
Manual
Cycle Time
79