Siemens SIRECUST 1280 Service Manual page 118

Bedside monitors
Table of Contents

Advertisement

SECTION
6.7
START
A
ANY
DISPLAY?
POWER
UP
OKAY
SOMETIMES,
BUT
SOMETIMES
NOT?
10
| АБ.
ADJUST
+65V
AND
+15V
5
NO
POWER
GREEN
LED
SUPPLIES.
REPLACE
INPUT
MODULE
BLINK
WHEN
GREEN
LED
ON?
POWER
SUPPLY
BOARD
FUSES
OKAY?
POWER
APPLIED?
(6.7A.1)
IF
PROBLEM
IS
RE
(6.74.3)
(6.7A.2)
NOT
CORRECTED.
1)
(6.7B.1)
a
VERIFY
THAT
HARNESSES
CHECK
POWER
OPERATOR
YES
INTERCONNECTING
INPUT
MODULE,
INTERFACE
LEDS
DISPLAY
SUBSECTION
REPLACE
IF
BAD.
TOGGLE
FOR
KEY
BOARDS
ARE
SECURELY
(6.7A.4)
PRESS?
CONNECTED
TO
(6.7A.17)
BOARDS.
(6.7A.18)
VERIFY
THAT
SOCKET
CHECK
THAT
BOARDS
POWER
YES
CONNECTOR
X4
ON
ARE
SECURELY
SEATED
SUPPLY
BOARD
POWER
SUPPLY
BOARD
ON
MOTHER
BOARD.
CHECK
CURVE
BOARD,
FUSES
OKAY?
Is
CONFIGURED
IF
PROBLEM
" REPLACE
IF
BAD.
(6.7A.5)
PROPERLY
FOR
LINE
IS
NOT
CORRECTED,
(6.7A.19)
VOLTAGE
AT
SITE.
CONTINUE.
(6.7A.7)
(6.7A.17)
REMOVE
LOADS
FROM
CHECK
SUPPLY
REPLACE
BLOWN
POWER
SUPPLY
BOARD.
VOLTAGES
FOR
FUSE(S).
IF
FUSE(S)
CHECK
SUPPLY
DISPLAY
SUBSECTION.
BLOW
AGAIN,
REPLACE
VOLTAGES.
REPLACE
REPLACE
POWER
POWER
SUPPLY
BOARD.
POWER
SUPPLY
BOARD
SUPPLY
BOARD
(6.7A.6)
IF
BAD.
IF
BAD.
(6.7A.8)
(6.7A.20)
|
x
CHECK
CONNECTION
REPLACE
FUSE(S).
IF
BETWEEN
POWER
FUSE(S)
BLOW
AGAIN,
PERFORM
SUPPLY
BOARD
AND
CHECK
POWER
INPUT
BOARD-BY-BOARD
TURN
Z-AMP
BOARD.
MODULE
AND
POWER
ON
PROCEDURE
TO
REPLACE
HARNESS
IF
SUPPLY
BOARD.
ISOLATE
BAD
BOARD.
BAD.
IF
PROBLEM
IS
(6.7A.3)
(6.7A.9
-
6.7A.16)
NOT
CORRECTED,
REPLACE
Z-AMP
BOARD.
(6.7A.21)
1
|
2
|
3
|
4
|
5
6
|
7
|
8
|
9
10
Figure 6-1.
Troubleshooting Flow-chart, Sheet 1 of 5