Testing Performance
To test the time interval accuracy (logic analyzer)

Connect the logic analyzer

Using a 6-by-2 test connector, connect channel 0 of Pod 1 to the pulse generator
1
channel 1 output.
Using the SMA cable and the BNC adapter, connect the External Input of the pulse
2
generator to the Main Signal of the function generator.

Acquire the data

Enable the pulse generator channel 1 output (with the LED off).
1
Press the blue key, then press the Run key to select Run-Repetitive. Allow the logic
2
analyzer to acquire data for at least 100 valid runs as indicated in the pattern
statistics field. Observe the X to O time field and ensure the X marker to O marker
time is between 47.496 and 47.504 µs during the test.
When the logic analyzer has acquired more than 100 valid runs, press Stop. The Min
3
X-O field, Max X-O field, and the Avg X-O field in the logic analyzer Pattern Statistics
menu should read 47.50 µs. Record the results in the performance test record.