Quality And Safety Checks - Miele PG 8582 CD Operating Instructions Manual

Washer-disinfector
Hide thumbs Also See for PG 8582 CD:
Table of Contents

Advertisement

Quality and safety checks

Thermo-electric
temperature
checks
Calibration of
dispensing
systems
Electrical safety
154
Factory tests
Every Miele cleaning machine undergoes extensive quality and safety
checks during the production process. They include the following
specific safety checks.
Thermo-electric temperature checks according to EN ISO 15883 incl.
disinfection parameters are carried out at the production plant.
Thermo-electric temperature checks do not have to be carried out
again during the initial commissioning of new cleaning machines.
Thermo-electric temperature checks are a mandatory requirement if
disinfection parameters (e.g. temperature, holding time, A0 value) are
changed during initial commissioning.
Thermo-electric temperature checks must be carried out in the
context of Operation Qualification (OQ) as part of performance
qualification according to EN ISO 15883.
Thermo-electric temperature checks must be carried out when a
machine is put back into operation after a period of downtime or
having been relocated, for example.
Regional and national rules and regulations must be complied with.
Calibration of dispensing systems according to EN ISO 15883 is
carried out at the production plant. Calibration of dispensing systems
can be omitted during the initial commissioning of new cleaning
machines.
Calibration of dispensing systems must be carried out in the context
of Operation Qualification (OQ) as part of performance qualification
according to EN ISO 15883.
Calibration of dispensing systems must be carried out when a
cleaning machine is put back into operation after a period of
downtime or having been relocated, for example.
Regional and national rules and regulations must be complied with.
Earthing and high-voltage testing according to IEC 61010-2-40 is
carried out at the factory.
If electrical installation and/or repair work proves necessary during
commissioning, an electrical safety check compliant with national
rules and regulations must be carried out.