Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
Exception
Error
ID
Warning
0405508
Warning
0405509
Error
0405510
Error
0405701
Error
0405702
Error
0405703
Error
0405704
Warning
0405705
Error
0405706
Error
6-54
Exception
Description
(Displayed in User Interface)
(In system Log)
SCU performs
DICOM storage failure. Please
storage
make sure the network and
operation failed
storage server configuration
are correct, and the server is
available. And try again. If this
issue persists, please call for
service.
SCU sets locale
System start failure. Set locale
failed
failed. Please restart the
system. If this issue persists,
please call for service.
SCU reports
System start failure. Please
ready failed
restart the system. If this issue
persists, please call for service.
SCU module
Workstation communication
resolve object
exception. Please restart the
exception
system. If this issue persists,
please call for service
SCU module
Workstation communication
Invoke corba
exception. Please restart the
service
system. If this issue persists,
exception
please call for service
SCU creates
Execution thread exception.
thread failed
Please restart the system. If
this issue persists, please call
for service.
SCU cancels
Cancel execution thread
thread failed
exception. Please restart the
system. If this issue persists,
please call for service.
SCU allocates
System run out of memory.
memory failed
Please restart the system. If
this issue persists, please call
for service.
SCU module
Workstation communication
register servant
exception. Please restart the
exception
system. If this issue persists,
please call for service
Message
Diagnosis Reference
Check DICM configuration.
System unknown error
make sure if omniNames
service failed, any module
crashed
make sure if omniNames
service failed, any module
crashed
make sure if omniNames
service failed, any module
crashed
System unknown error
System unknown error
Check memory using state
make sure if omniNames
service failed, any module
initialization failed
Need help?
Do you have a question about the Brivo OEC 715 and is the answer not in the manual?
Questions and answers