To help ensure operators are adequately trained in the safe use of the equipment, STERIS recommends initial training and periodic proficiency training of all personnel concerned with the operation and maintenance of the equipment; that attendance records be maintained;...
1 — A Word From STERIS 1.5 Intended Use and Clinical Benefit The AMSCO 600 Steam Sterilizer is intended for sterilization of heat and moisture-stable materials used in healthcare facilities. The sterilizers are generally operated by technicians in a central service or sterile processing department of healthcare facilities.
1 — A Word From STERIS 1.6 Technical Specifications AMSCO 600 Sterilizer design parameters include: Table 1-2. Dimensions Arrangement External Dimensions (Depth x Width x Height) (mm/inch) Model 639V-3 Model 851V-3 Model 1063V-3 1464 x 1197 x 2261 mm 1764 x 1197 x 2261 mm...
TABLE OF CONTENTS Section Number Description Page A Word From STERIS............................. 2 Copyright and Publish Date ....................... 2 Service Information ........................2 Advisory............................2 Sterilizer Safety Devices......................2 Intended Use and Clinical Benefit....................3 Technical Specifications ......................4 Associated Publications ......................4 Environmental Conditions......................
Page 8
Routine Maintenance ..........................64 Preventive Maintenance ......................64 Clean Chamber Drain Strainer (Daily) ..................65 Routine Maintenance Procedure ..................... 65 Thermal Printer......................... 69 Recommended Spare Parts ..................... 71 Troubleshooting............................72 Troubleshooting – General Information ................. 72 Air Detected in Drain - ALM ..................... 74 Appearance Issues........................
Page 9
6.43 Generator – Water Leak......................88 6.44 Generator – Water Level Control Issue................... 89 6.45 Generator CSG – Control Power is OFF .................. 89 6.46 Generator CSG – Failure to Build Pressure................89 6.47 Generator CSG – No Water in Sight Glass - Pump Does Not Start ........90 6.48 Generator CSG –...
Page 10
6.92 Too Long To Close NOE Door - ALM ..................105 6.93 Too Long To Close OE Door - ALM ..................106 6.94 Too Long to Depressurize Generator ..................106 6.95 Too Long To Evacuate ......................107 6.96 Too Long To Open Ball Valve-ALM..................107 6.97 Too Long to Open NOE Door....................108 6.98...
SURFACE CHAUDE GLISSANTE NOTES: WARNING TOLERANCE STANDARD FRACTIONAL ±1/64 DECIMAL ± .005 1. ARTWORK TO BE SUPPLIED BY STERIS CONCURRENT ENGINEERING. UNLESS OTHERWISE NOTED ANGULARITY ±1° MACH. SURF. 2. MATERIAL SPECIFICATIONS AND MANUFACTURER TO BE IN PRECAUCION STERIS Corporation TITLE: Mentor, OH ACCORDANCE WITH STERIS DEWG.
Page 12
2 — Safety Precautions Table 2-2. Icons Icon Definition Icon Definition Machine Setup Sound Options Change Cycle Values Start Cycle Standby Advance Screen Exit Screen (back one Options screen) Abort Cycle OK (Accept) Backspace(Erases entered Display Abort/Alarm Status characters) Set Print Options Alarm Details Set Machine ID Silence Alarm...
IMPORTANT: Read this document to get the best understanding and to help ensure best performance of this equipment, which has been thoroughly inspected, tested and adjusted by factory-trained personnel. Equipment operation or maintenance other than described in the manual (unless by authorized representatives of the STERIS Corporation), shall be at the risk of invalidating the equipment warranty.
Page 14
The performance of the sterilizer is validated as a system including components defined by STERIS in the Operator and Maintenance Manuals for the sterilizer. Substitution of unauthorized components can potentially lead to personal injury, damage or premature failure of the product and result in a unit configuration that is inconsistent with the validated product.
Page 15
2 — Safety Precautions Repairs and adjustments to this equipment must be made only by STERIS or STERIS-trained personnel. Maintenance performed by unqualified personnel or installation of unauthorized parts could cause personal injury, result in improper equipment performance, void the warranty or result in costly damage.
Page 16
2 — Safety Precautions WARNING SLIPPING HAZARD To reduce slippery conditions, keep floors dry by immediately wiping up any spilled liquids or condensation in sterilizer loading or unloading areas. WARNING STERILITY ASSURANCE HAZARD Load sterility may be compromised if the Bowie-Dick test or air leak test indicates a potential problem. If these indicators show a potential problem, refer the situation to a qualified service technician before using the sterilizer further.
Page 17
2 — Safety Precautions Removing the door seal gasket with a tool, such as a screwdriver, is NOT recommended as, in most cases, it will cut or damage the gasket so that it cannot be further used. Do not remove the gasket unless absolutely necessary.
3 — Component Identification WARNING ELECTRIC SHOCK AND BURN HAZARD Do not use control box shutoff switch to isolate power. The switch does not turn off single-phase power. Shutoff all utilities to sterilizer before servicing. Do not service the sterilizer unless all utilities have been properly locked out.
3 — Component Identification Figure 3-3. Control Panel Components, OE View 1. Panel Release 2. Chamber Pressure Gauge 3. Jacket Pressure Gauge 4. Emergency Stop 5. Thermal Printer 6. USB Port 7. Touch Screen 3.3 Control Panels - Double Door Units Only Double-Door units have two control panels;...
3 — Component Identification 3.5 Sterilizer Touch Screen Control Touch-screens respond to very light pressure. The sterilizer switches from In Standby to Ready mode when the screen is touched. In Standby mode is usually used at night when the sterilizer is not being operated and steam is turned off to save energy.
3 — Component Identification Figure 3-7. Sliding Door Operation 3.7 Operating End and Non-Operating End Door Control Panels A sterilizer equipped with two doors is also equipped with two control panels. The control panel at the loading door of the sterilizer is referred to as Operating End (OE) Door Control; the control panel located at the unloading door is referred to as Non-Operating End (NOE) Door Control.
3 — Component Identification To open a panel, use the included key to unlock the associated locks for a given panel, then carefully remove the panel from the sterilizer. After reinstalling a panel, be sure to lock all associated compression locks to prevent accidental opening or dislodging. 3.9 Pressure Gauges The front (OE) panel includes two front dial pressure gauges, located above the chamber.
3 — Component Identification 3.11 Printouts The printout reports useful information about each cycle the sterilizer runs; including the load number, which is a unique identifying code. Each load number is printed in this numeric format: 022321-01001 Understanding the load number: The example load number of 022321-01001 indicates: February 23, 2021, Sterilizer No 1, first cycle.
3 — Component Identification 3.12 USB Ports (for Cycle Storage / Cycle Reprint) Sterilizer stores up to the last 300 cycles on internal memory; these internally stored cycles can be retrieved, viewed, printed or copied to a USB stick. The sterilizer includes a USB port on the OE panels. The USB port is located below the E-Stop on the NOE side, by default, on double door units.
Page 28
3 — Component Identification Emergency Stop Key An emergency stop key is required to reset the emergency stop switch. NOTE: The emergency stop key should be retained by the department supervisor. Resetting the Switch: 1. Insert emergency stop key. 2. Turn emergency stop key counterclockwise. a.
4 — Operating Instructions Operating Instructions 4.1 General Concepts is important to realize the following general concepts: • Prior to sterilization, all materials and articles must be thoroughly cleaned. • After sterilization, store all items as to not compromise the barrier quality of their wrapping materials. •...
4 — Operating Instructions 1. Press anywhere on the sterilizer touch screen. Printer records the time and date that the power is turned on. 2. Open chamber door using Door Open/Close button. a. Check that drain strainer is clean and in place and that chamber interior is clean. See Section 5.2, Clean Chamber Drain Strainer (Daily), if cleaning is necessary.
4 — Operating Instructions • Place items on shelves such that the heaviest items reside on the bottom shelf and the lighter items reside on the higher shelf. • In mixed loads of fabrics and hard goods, place the hard goods on lower shelf. This placement reduces wetting of fabric packs which would result from condensate dripping from hard goods.
Page 33
4 — Operating Instructions Loading WARNING BURN HAZARD Sterilizer, rack/shelves, and loading car will be hot after cycle is run. Always wear Personal Protective Equipment (PPE) when removing a processed load. PPE must also be worn when reloading sterilizer following previous operation. Do not load transfer carriage, loading cars or rack and shelf assembly beyond specified limits.
Page 34
4 — Operating Instructions IMPORTANT: After removing load from the chamber, close the chamber door and keep the chamber door closed to minimize utility consumption. NOTE: Water droplets and visible signs of moisture on sterile packaging or the tape used to secure it, may compromise sterility of processed loads or be indicative of a sterilization process failure.
4 — Operating Instructions Figure 4-2. Rack and Shelves Dimensions 4.6 Loading Car and Transfer Carriage Loading 1. Open sterilizer door using Door Open/Close button on touch screen. IMPORTANT: Ensure loading cars are never loaded beyond their specified load limit (see Table 1-3).
Page 36
4 — Operating Instructions WARNING PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD ● Do not load transfer carriage, loading cars or rack and shelf assembly beyond validated limits. ● Ensure transfer carriage is locked to chamber rails before pushing loading car into chamber. 4.
4 — Operating Instructions 4.7 Unloading Car and Transfer Carriage WARNING BURN HAZARD ● Sterilizer, rack/shelves and loading car will be hot after cycle is run. Always wear Personal Protective Equipment (PPE) when removing a processed load. PPE must also be worn when reloading sterilizer following previous operation.
4 — Operating Instructions NOTE: The Latch and Release Lever secures the loading car to the transfer carriage while simultaneously unlocking and releasing the transfer carriage from the sterilizer chamber rails. 6. Carefully back the transfer carriage away from sterilizer. 7.
4 — Operating Instructions Table 4-1. Interlock Codes and Functionality (continued) Comments Concept Image Code Door Door • At end of cycle operator must manually unseal the unload door by pressing OPEN DOOR button on the touch screen. Unload Load •...
4 — Operating Instructions IMPORTANT: If desired cycle parameters (sterilize temperature, sterilize time, dry time) are other than those listed in the operator manual, it is the responsibility of the Customer’s facility to consult and follow the device manufacturer’s written instructions regarding other parameters.
Page 41
4 — Operating Instructions WARNING PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD When closing the chamber door, keep hands and arms out of the door opening and make sure opening is clear of obstructions. Before running this cycle refer to Section 4.1, General Concepts.
Page 42
4 — Operating Instructions NOTE: The touch screen indicates cycle progress in three ways (refer to Section 4.9, Introduction to Sterilizer Cycles for examples of screens): a. Digital timer (in screen center) shows estimated remaining cycle time. b. A change of phase name in the top status line of the screen. c.
4 — Operating Instructions 4.11 Aborting Cycles It may be necessary to end a processing cycle, possibly because the wrong cycle was selected or the sterilizer begins functioning incorrectly. A cycle can be aborted at any time by pressing the ABORT button. WARNING BURN HAZARD Sterilizer, rack/shelves and loading equipment will be hot after cycle is run.
4 — Operating Instructions 4.12 Cycle Selection and Change Values Cycle Selection Changing Cycle Values Aborting a Cycle Cycle Selection AMSCO® 600 Series Steam Sterilizer is shipped with factory-set cycles, cycle values and control values programmed into the control. IMPORTANT: If cycle parameters (sterilize time, dry time, temperature) are not as desired by the user, it is the responsibility of the Customer’s facility to consult and follow the device manufacturer’s written instructions.
Page 46
4 — Operating Instructions Changing Cycle Values As discussed in the Supervisor topic, Change Values Cycle Selection touch-screen options are accessed through the Supervisor Menu screen by pressing CHANGE CYCLE VALUES. This screen is used for changing cycle operating values, changing the cycles displayed (on the cycle selection menus) and for changing the operating settings of the sterilizer.
4 — Operating Instructions 4.13 Options Menu General information Machine Status Print Last Cycle Sound Options Supervisor Leaving the Options Menu General information The options menu allows the operator to access the machine status, print a cycle tape of the last cycle run, change sound tones and end of cycle sound duration, and access the supervisor menu.
Page 48
4 — Operating Instructions Print Last Cycle If a duplicate print of the last cycle run is required, access the Options menu and press Duplicate Print to produce a second printout. Sound Options The operator can adjust selected sounds and tones heard at the sterilizer control and set the sound duration at the end of a cycle.
4 — Operating Instructions • Touch Tone is the sound heard whenever a button is pressed. Choices are Low, Medium, High, Off. • End of Cycle Sound is the sound heard when a cycle is complete. Choices are Low, Medium, High, Off. •...
Page 50
4 — Operating Instructions The supervisor menu is used to adjust sterilizer operation and includes a password feature. If the Supervisor Password is enabled, all options are secured for use by the supervisor only (see Change Password for details). NOTE: Enter password, if control is password enabled, to access Supervisor screen. Use arrow buttons to switch from first to second Supervisor screen and back, as needed.
Page 51
4 — Operating Instructions b. Set Month i. At the Set Date screen, press the Set Month button and the control advances to the Set Month screen. ii. Enter correct month value using the keypad. Month value entered must be between 1 and 12. iii.
Page 52
4 — Operating Instructions • MM-DD-YY or MMM-DD-YY • DD-MM-YY or DD-MMM-YY • YY-MM-DD or YY-MMM-DD iii. Once the appropriate format has been selected, or if format does not need to be changed, press back arrow button to return to Options screen. 5.
Page 53
4 — Operating Instructions i. If an incorrect number is entered, press backspace key to clear entry, then use keypad to enter correct value. ii. Once correct time has been entered, press OK to accept and save change hours screen. iii.
Page 54
4 — Operating Instructions i. If an incorrect value is entered, press the backspace key to clear the entry, then use the keypad to enter the correct value. ii. Once the correct minutes value has been entered, press OK to accept and save. iii.
Page 55
4 — Operating Instructions 2. The Password entry screen is displayed. • Use the keypad to enter the new password, then press OK. Use the backspace button to erase unwanted characters. • To disable the password, enter zero, then press OK. The Supervisor screen is displayed. Cycle Setup Cycle values are adjustable within the limits shown in the table.
Page 56
4 — Operating Instructions Table 4-2. Cycle Parameters, Adjustable by Supervisor (continued) CYCLE NAME STERILIZE STERILIZE TIME (hh: DRY TIME (hh:mm:ss) TEMPERATURE (°C/° mm:ss) Warm-Up 134°C/273°F / Not 03:00:00 / Not 00:01:00 / Not Adjustable Adjustable Adjustable Air Detector Test 00:01:00 / Not Adjustable NOTE: Interlock option (selections 0 - 6) are available for all cycles.
Page 57
4 — Operating Instructions 7. Press one of the displayed cycles, or press the Next arrow to display cycles not currently shown. Once the cycle has been found and selected, the control displays basic cycle information. NOTE: Test Cycle parameters cannot be changed so are not available for selection. After selecting and pressing the cycle to be changed, the control advances to a set of Cycle Parameters screens.
Page 58
4 — Operating Instructions NOTE: The control advances to the set up screen for the selected parameter so it can be set to a value within a predefined range as indicated in Table 4-2. 9. Once all changes have been made, press the back arrow button once or twice, as needed, to exit the Cycle Parameters screen(s).
Page 59
4 — Operating Instructions Table 4-3. Machine Setup Details (continued) Print Format–allows for switching between full printing format or paper-saving condensed printout format. Auto Duplicate Print–Provides a means of having a duplicate printout created automatically at the end of each cycle. The default is for single printout (no duplicate print).
Page 60
4 — Operating Instructions Utility Control (J) Utility Control screen is used to automatically control sterilizer steam supply and, for units with integral steam generators, set the time for automatic flush. 1. To access Utility Control, navigate to the Machine Setup screen. 2.
Page 61
4 — Operating Instructions – Manual – Off All Day – Flush (toggle) – flush option for each day has a toggle to flush [or not flush] and set flush time. • Startup Time – set time for sterilizer utilities startup. Upper portion of this button will show N/A unless the Function button was used to "Set Time."...
Page 62
4 — Operating Instructions 3. Select the desired function (Enable Printer, Print Format, Auto Duplicate Print). Enable Printer – This option allows the Customer to turn the printer on or off. Note that if the printer has been Disabled, print data is still retained by the sterilizer control and can be accessed at Cycle Print Log. 1.
Page 63
4 — Operating Instructions 4. If an incorrect character is entered, press the backspace key (not to be confused with the back arrow key) to clear the entry, then use the keypad to enter the correct value. 5. Press OK to save Customer Name or press back arrow button to abandon changes and return to the Machine Setup screen.
Page 64
4 — Operating Instructions From the Set Reminder screen you can: • Cycle Log Limit Warning Enabled? - Set to YES or NO • Weekly Leak Test Reminder - Set to ENABLED or DISABLED • Daily Bowie-Dick Test Reminder - Set to ENABLED or DISABLED •...
4 — Operating Instructions 6. Print (or save file to USB stick) by pressing either the print button (upper left) or save-to-disc button (lower right), as applicable If saving data, first install a USB stick into a USB port on the sterilizer. USB port location is just above the Emergency Stop Button.
Page 66
4 — Operating Instructions This cycle is used to warm sterilizer shelves, door and jacket before a Bowie-Dick cycle is run. NOTE: Always run this test cycle with either racks installed, or with an empty car in the chamber. 1. Refer to Section 4.1, General Concepts, and to Section 4.3, Before Operating the...
Page 67
4 — Operating Instructions NOTE: Report any indication of malfunction to the supervisor. The sterilizer must not be used to run prevacuum cycles until approved by supervisor. Bowie-Dick Test cycles do not test for adequate exposure to heat in terms of time-at-temperature. WARNING BURN HAZARD Sterilizer, rack/shelves and loading equipment will be hot after cycle is run.
Page 68
4 — Operating Instructions 7. Start Bowie-Dick/Air Removal test cycle. Cycle runs automatically after pressing the associated as described in the bulleted phase list in Section 4.10, Prevac Cycles. WARNING BURN HAZARD Sterilizer, rack/shelves and loading equipment will be hot after cycle is run. Always wear Personal Protective Equipment (PPE) when removing a processed load.
Page 69
4 — Operating Instructions Figure 4-13. Bowie-Dick Cycle Tape Vacuum Leak Test This cycle is used for testing vacuum integrity of the sterilizer and piping. IMPORTANT: The Vacuum Leak Test is not a substitute for a Bowie-Dick test. 14159673_A Operator Manual AMSCO®...
Page 70
4 — Operating Instructions WARNING BURN HAZARD ● Sterilizer, rack/shelves and loading equipment will be hot after cycle is run. Always wear Personal Protective Equipment (PPE) when removing a processed load. PPE must also be worn when reloading sterilizer following previous operation. ●...
Page 71
(a different test normally performed during operational qualification). During verification of the sterilizer, STERIS adjusted this valve to give a leak of 10 mbar/min (as measured during a normal Vacuum Leak Test with this valve manually forced open on the air manifold. This valve is labeled “S99” on the air manifold. For a picture of the air manifold, refer to Section 7.1, Emergency/Manual Door Opening...
Page 72
4 — Operating Instructions Figure 4-15. Location of Air Detector Metering Valve 4. Load a standard 7 kg textile test pack (see EN 285 23.1 Standard test pack) on the bottom shelf over the chamber drain. 5. Start Air Detector test cycle. Cycle runs automatically after pressing the associated as described in the bulleted phase list in Section 4.10, Prevac Cycles.
Regularly scheduled preventive maintenance is required for safe and reliable operation of this equipment. Contact your STERIS Service Representative to schedule preventive maintenance. The performance of the sterilizer is validated as a system including components defined by STERIS in the Operator Manual and Service Manual for the sterilizer. Substitution of unauthorized components can potentially lead to personal injury, damage or premature failure of the product and result in a unit configuration that is inconsistent with the validated product.
5 — Routine Maintenance NOTE: Preventive Maintenance is not covered under warranty. Follow all safety procedures including Lockout/Tagout. Inspection Frequency: The sterilizer must be inspected by a qualified service technician at least twice each year while referring to the MAINTENANCE MANUAL (actual frequency of equipment use and local conditions may increase the number of required inspections).
Page 76
NOTE: For detailed chamber cleaning instructions, refer to AMSCO Steam Sterilization Cleaning Procedure M2826, available from STERIS. The detergent recommended in this cleaning procedure may not be available in your area. If it is the case, contact your STERIS representative for alternative solutions.
Page 77
5 — Routine Maintenance WARNING BURN HAZARD Allow sterilizer to cool to room temperature before performing chamber cleaning. 1. With sterilizer shutdown, cool, and door open, perform proper LOCK-OUT/TAG-OUT procedures (per local regulations) for facility steam, electric, and water utilities which supply the sterilizer. Figure 5-3.
Page 78
5 — Routine Maintenance NOTE: The shelving rack can only be removed by a trained technician. Refer full chamber cleaning to STERIS- trained personnel. Figure 5-5. Remove Track Assembly for Proper Chamber Cleaning CAUTION POSSIBLE EQUIPMENT DAMAGE Never use a wire brush, abrasives, or steel wool on door and chamber assembly. Do not use cleaners containing chloride on stainless-steel surfaces.
Cleaners, when rubbed in a circular motion or applied with a wire brush or steel wool, can damage stainless-steel door and chamber assemblies. To restore the original luster of the stainless-steel surfaces STERIS recommends using Stainless-Steel Cleaner, STERIS product number NM 410. Refer to Safety Data Sheet SDS A301 for detailed product information.
Page 80
5 — Routine Maintenance Figure 5-6. Thermal Printer 2. Lift Inner Cover as shown in photo. 3. Position paper roll as follows: a. Position paper roll in bottom of printer compartment, paper coming from bottom of roll. b. Ensure at least 8” (203 mm) of paper extends from printer compartment. 4.
8. Close printer cover. 5.5 Recommended Spare Parts NOTE: Use only STERIS authorized parts on this equipment. Use of unauthorized parts will void the warranty. The following replacement parts are recommended to be kept on hand: • Paper, Printer, Thermal — part number: 10092606 (3 rolls) •...
Allow sterilizer to cool to room temperature before performing any cleaning or maintenance procedures. If a problem occurs that is not described in this section, please contact STERIS Corporation. A trained service technician will promptly place your sterilizer in proper working condition.
Page 83
6 — Troubleshooting Figure 6-1. Example Alarm, Alarm Details, and Alarm Service Screens Legend: 1. Press here to get alarm details (2nd screen) 2. Silence Alarm 3. Chamber Status 4. Acknowledge Alarm 5. Press here to get service details (3rd screen) NOTE: The 'Cycles prevented, please call Service' screen appears when an alarm occurs that will prevent any cycle from being started until Service is called.
RTD8, reads below the configurable “Air Detect Alarm” service parameter at the end of the first pressure pulse in the Prevac conditioning phase. General Troubleshooting Additional Information Internal malfunction 1. Silence alarm. User must acknowledge alarm to clear screen. 2. Contact STERIS. 14159673_A AMSCO® 600 Medium Operator Manual...
General Troubleshooting Additional Information Dirty or mis-adjusted fascia panels. 1. Clean or adjust fascia panels if needed. 2. Contact STERIS. 6.4 Barcode Scanner Error Condition NOTE: Applies only to units with Automated Load/Unload System (ALUS) installed. Sterilizer will not automatically load or unload the chamber. Display indicates Barcode Scanner Error. Printer prints Barcode Scanner Error.
If door operates normally, the machine can be operated until STERIS Service arrives. b. Contact STERIS and report condition. 2. If door is completely open, do not operate machine and contact STERIS to report the condition. 14159673_A AMSCO® 600 Medium Operator Manual...
1. Reset the control by powering down the sterilizer using the power switch on the sterilizer control box. 2. If condition recurs, contact STERIS. Connection issue. 1. Request biomed to confirm unit has proper 3 phase supply voltage to sterilizer.
Door A open 1. Acknowledge alarm. 2. Attempt to close door. 3. If alarm recurs, acknowledge it and contact STERIS. 6.12 Door A Switch Failure-ALM Condition OE Door Switch Failure alarm is displayed. Sterilizer has aborted the cycle. Printer prints OE Door Switch Failure.
The Door B Open Alarm shall occur if the NOE door closed input is sensed OPEN while the door is sealed by the software. General Troubleshooting Additional Information Silence alarm. 1. Acknowledge alarm. 2. Attempt to close door. 3. If alarm recurs, acknowledge it and contact STERIS. 14159673_A Operator Manual AMSCO® 600 Medium...
General Troubleshooting Additional Information Silence alarm 1. User must acknowledge alarm to clear screen. 2. Contact STERIS. 6.17 Door B Unsealed - ALM Condition NOE Door Unsealed alarm is displayed. The sterilizer has aborted the cycle. Printer prints NOE Door Unsealed.
2. Use e-stop key and rotate key until switch pops out. 3. If alarm recurs, call STERIS. 6.21 E Stop A Released Alarm Condition Emergency Stop Released OE Door alarm is displayed. Indicates that the e-stop has changed state from pressed to released.
2. Use e-stop key and rotate key until switch pops out. 3. If alarm recurs, call STERIS. 6.23 E Stop B Released Alarm Condition Emergency Stop Released NOE Door alarm is displayed. Indicates that the e-stop has changed state from pressed to released.
Additional Information Silence alarm 1. User must acknowledge alarm to clear screen. 2. Call STERIS 6.27 Failure Reading NOE Door Pressure - ALM Condition Display reads Failure reading NOE Door Pressure. Cycles are prevented until serviced. Printer prints Failure Reading PT4.
General Troubleshooting Additional Information Internal malfunction 1. Silence alarm. 2. Contact STERIS. 6.30 Failure Reading Record Temp-ALM Condition Display reads Failure Reading Recorder Temperature. Printer prints Failure Reading Record T. Alarm occurs immediately when the system encounters a failure in reading the analog signal for the recorder temperature.
Additional Information Acknowledge alarm. 1. While no cycle is running, power down the sterilizer and re-start. 2. If alarm recurs, call STERIS. 6.33 GDC Switch Failure-ALM Condition Display reads GDC Switch Failure. Printer prints GDC Switch Failure. Generator flush is aborted and cycles will be prevented until serviced.
Additional Information Control voltage not present 1. Contact facility maintenance to verify generator control voltage. 2. Contact STERIS. 6.36 Generator – Failure to Build Pressure Condition Too Long to Charge Jacket alarm or Too Long To Charge alarm may be displayed.
Pump does not start. 1. Contact facility maintenance to verify single phase generator control voltage. 2. Contact STERIS if voltage present. 6.39 Generator – Pump Noisy Condition Unusual noise from sterilizer while generator water pump is running. Cycles are not affected.
1. Evaluate if there is a safety concern where water is found. If yes, complete any cycle and shut down the sterilizer. Contact STERIS. 2. If puddled water is not a safety concern, contact STERIS as soon as possible. 14159673_A AMSCO® 600 Medium Operator Manual...
2. Contact facilities to confirm generator water pressure is between 137.9 and 344.7 kPa (1379 and 3447 mbar). 3. Contact STERIS. 6.45 Generator CSG – Control Power is OFF Condition Display indicates Too Long to Charge Jacket or Too Long to Charge alarm.
Pump does not start. 1. Contact facility maintenance to verify there is approximately 230V going to generator control box. 2. Contact STERIS if voltage present. 6.48 Generator CSG – Pump Noisy Condition Unusual noise from sterilizer while generator water pump is running. Cycles are not affected.
1. Evaluate if there is a safety concern where water is found. If yes, complete any cycle and shut down the sterilizer. Contact STERIS. 2. If puddled water is not a safety concern, contact STERIS as soon as possible. 14159673_A Operator Manual AMSCO® 600 Medium...
2. Contact facilities to confirm generator water pressure is between 137.9 and 344.7 kPa (1379 - 3447 mbar). 3. Contact STERIS. 6.55 Intermittent Valve Operation Condition Sterilizer is exhibiting random, inconsistent alarms or overall performance over time. Possible intermittent wet loads observed.
Intermittent door alarms occur. General Troubleshooting Additional Information Acknowledge any alarm. 1. Note conditions that alarms occur for Service Technician. 2. Call STERIS 6.58 Leaks - Air line Condition Hissing noise coming from sterilizer. General Troubleshooting Additional Information Air line cracked, press on connections faulty, air lines not 1.
2. Run a second leak rate test. a. If the second test passes, the machine can be used. b. If the second test fails, contact STERIS. Leak test run with cart in chamber causing condensate to 1. Remove cart and rerun test.
General Troubleshooting Additional Information Internal issue 1. Silence alarm. Do not power down the sterilizer. 2. Contact STERIS. 6.65 Memory Write Error - ALM Control Issue Condition Alarm when an error occurs when writing to non-volatile memory. Display indicates memory Write Error. Printer prints Memory Write Error. All functions are stopped and cycles cannot be run until serviced.
The door is obstructed 1. Silence alarm. 2. Clear obstructions. 3. User must acknowledge alarm to clear screen. 4. If alarm recurs, contact STERIS. Lower access panel air supply cutoff switch is not 1. Lock lower door panel. engaged. 6.68 NOE Door Open Switch Failure - ALM Condition Display indicates NOE Door Open Switch Failure.
2. Identify if noise is affecting processing loads. If yes, shut down the sterilizer and contact STERIS. 3. If noise is not affecting loads, the sterilizer may continue to be used until STERIS Service can be contacted. 6.73 Noise — Water Hammer Condition Periodic loud banging noise is heard from sterilizer area.
Silence alarm 1. Clear obstructions 2. User must acknowledge alarm to clear screen. 3. If alarm recurs, call STERIS 6.76 OE Door Open Switch Failure - ALM Condition Display indicates OE Door Open Switch Failure. Printer prints Door Switch Failure OPNA.
551.6 - 689.5 kPa (5516 - 6895 mbar). 3. Run cycle. 4. Contact STERIS if issue recurs. 6.78 Over Sterilize Temperature - ALM Condition Display indicates Over Sterilize Temperature. Printer prints Over Sterilize Temp. The sterilizer will abort the current action including a running cycle.
2. Verify with facility maintenance that power has been restored. 3. Restart the cycle following facility SOPs. 4. Contact STERIS if sterilizer fails to start. 6.80 Pressure In Chamber - ALM Condition Display will indicate Pressure In Chamber. Printer will print Pressure In Chamber.
6 — Troubleshooting 6.81 Printer Failure-ALM Condition Display will indicate Printer Failure. Nothing will print. This is an immediate alarm that notifies the user of a printer failure. It does not prevent cycles from being run. The alarm is triggered when there is an error from the system indicating communication error with the printer. Sterilizer will continue operation without printing.
General Troubleshooting Additional Information Silence alarm 1. Recalibrate or reload default calibration values. 2. Call STERIS. 6.84 Real Time Clock Fail-ALM Condition Display indicates Real Time Clock Failure. Printer prints Real Time Clock Fail. Real Time Clock Failure is an immediate alarm which aborts the cycle and prevents additional cycles from running until service has resolved the problem.
General Troubleshooting Additional Information Acknowledge alarm 1. Rerun cycle 2. If alarm recurs, call STERIS. 6.88 Too Long In Charge – ALM Condition Display indicates Too Long in Charge. Printer prints Too Long to Charge. This alarm occurs when the system is in the charge step and the sterilize temperature has not been achieved within thirty minutes.
137.9 and 344.7 kPa (1379 and 3447 mbar). 3. Rerun cycle. 4. If alarm recurs, contact STERIS. 6.90 Too Long In Jacket Charge - ALM Condition Display will indicate Too Long to Charge Jacket. Printer will print Too Long to Chrg Jckt.
2. Later production sterilizers have a door safety interlock that will prevent the door from moving if not fully engaged. Confirm lower access door is fully locked if so equipped. 3. If door will still not close, contact STERIS. 14159673_A Operator Manual AMSCO® 600 Medium...
Confirm lower access door is fully locked if so equipped. 3. If door will still not close, contact STERIS. 6.94 Too Long to Depressurize Generator Condition Display indicates Too Long to Depressurize Generator. Printer will print Too Long to Depress Gen. Automated flush will not occur if equipped.
137.9 & 344.7 kPa (1379 & 3447 mbar). 3. Check water supply valve. If closed, open valve. 4. If alarm recurs, abort cycle and contact STERIS. 6.96 Too Long To Open Ball Valve-ALM Condition Display indicates Too Long to Open Ball Valve. Printer prints Too Long to Open BV.
2. Confirm lower access door is fully locked if so equipped (later production sterilizers have a door safety interlock that will prevent the door from moving if not fully engaged). 3. If door will still not open, contact STERIS. 14159673_A AMSCO® 600 Medium Operator Manual...
Too long to pressurize door 1. Silence alarm. User must acknowledge alarm to clear screen. 2. If alarm recurs, contact STERIS. 6.100 Too Long to Pressurize OE Door Condition Display indicates Too Long to Pressurize OE Door. Printer prints Too Long to Press OE Dr.
3. Correct as required. 4. Contact STERIS if alarm recurs. If electric steam generator is used - 3 phase voltage is 1. Contact facility biomed to ensure 3 phase voltage is unavailable available.
• Extend exhaust time General Troubleshooting Additional Information Silence Alarm 1. Call STERIS 6.104 Too Long To Unseal NOE Door - ALM Condition Display indicates Too Long to Unseal NOE Door. Printer prints Too Long to Unseal NOE Dr. Alarm occurs when the system takes longer than 300 seconds (5 minutes) to unseal NOE Door.
3. Correct as required. 4. Contact STERIS if alarm recurs. If electric steam generator is used - 3 phase voltage is 1. If electric steam generator is used, contact facility unavailable biomed to ensure 3 phase voltage is available.
3. Correct as required. 4. Contact STERIS if alarm recurs. If electric steam generator is used - 3 phase voltage is 1. If electric steam generator is used, contact facility unavailable biomed to ensure 3 phase voltage is available.
2. If the water is not deemed to be a safety issue, clean up the water spill, note the location of the spill and contact STERIS. Monitor for any further water collection and dry the area as needed until STERIS Service arrives.
7 — Appendix Appendix 7.1 Emergency/Manual Door Opening Procedure WARNING PERSONAL INJURY HAZARD AND/OR EQUIPMENT DAMAGE HAZARD Emergency Door Procedure is not a typical procedure and is only to be performed by properly trained personnel. Prior to performing the procedure review all WARNINGS and CAUTIONS associated with the sterilizer and review the facilities Lock-Out/Tag-Out procedures, which must meet or exceed local Standards.
Page 127
7 — Appendix IMPORTANT: The sterilizer may be equipped with a mechanical air switch that prevents door operation when the lower panel is removed. Make sure the lower panel is in place and securely locked. Features of the air switch: mechanical-only, no electrical or software association switch activates when lower panel lock is in the unlock position switch remains activated when lower panel is not in place...
Page 128
7 — Appendix Figure 7-3. Component Locations, Needed for Opening Chamber Door Legend: a. Electrical Power-In 3 Phase b. Power Box c. Control Box d. Control Box Shutoff Switch e. MV3 – Air Supply (Open) f. MV2 – Steam Supply (Open to Steam Generator) g.
Page 129
7 — Appendix WARNING BURN HAZARD Stay clear of areas surrounding operating end (OE) chamber door; steam and hot water may be released from chamber when OE door gasket unseals in step 3 or 4 below. 3. On pneumatic valve air manifold (E), press in and rotate upper detent ¼-turn clockwise to lock OE DOOR SEAL EXHAUST pneumatic valve (S37) OPEN;...
Need help?
Do you have a question about the AMSCO 600 Medium 639V-3 and is the answer not in the manual?
Questions and answers