2.2.4 How to Mount
5 mL Syringe
Mount the following parts (see Figures 1-3, 2-5, 2-6 and 2-7).
(1)
Mount the 5 mL syringe.
(2)
Using the spanner furnished together with the L-2200/
2200U autosampler, mount the sample loop (890-3464)
and connecting coil (890-3465).
Fig. 2-7 Mounting of 5 mL Syringe
NOTICE:
1. Install the connecting coil (890-3465) on the
cover of the autosampler.
Store the sample loop (890-3464) in the storage
space at the side of injection valve (see Figure 2-
6).
2. Push the sample loop (890-3464) completely
into the needle coupling, and then tighten
setscrew 1 (see Figure 2-5).
2 - 7
Connecting coil
(890-3465)
Syringe
(080-1593)
Sample loop
(890-3464)