2.2 Setup
2.2.3 How to Mount
0.5 mL/1 mL/

2.5 mL Syringe

Fig. 2-4 Mounting of 0.5 mL/1 mL/2.5 mL Syringe
Mount the following parts (see Figures 1-2, 2-4, 2-5 and 2.6)
(1)
Mount the 0.5 mL/1 mL/2.5 mL syringe.
(2)
Using the spanner furnished together with the L-2200/
2200U autosampler, mount the sample loop 1 (890-3481),
coupling, sample loop 2 (890-3466, 890-3468, 890-3470),
and connecting cil (890-3467, 890-3469, 890-3471).
NOTICE: 1. Push sample loop 1 (890-3481) completely into
the needle coupling, and then tighten setscrew 1
(see Figure 2-5).
2. Store sample loop 2 (890-3466, 890-3468, 890-
3470) and connecting coil (890-3467, 890-3469,
890-3471) in the storage space at the side of
injection valve (see Figure 2-6).
2 - 4
Syringe
(080-1064, 080-1592,
080-1092)
Sample loop 2
(890-3466, 890-3468,
890-3470)
Sample loop 1
(890-3481)
Connecting coil
(890-3467,
890-3469,
890-3471)
Coupling