Signal conditioning extensions for instrumentation (60 pages)
Summary of Contents for National Instruments NI 9155
Page 1
NI 9155. This document also contains specifications for the NI 9155. This document refers inclusively to all reconfigurable embedded chassis with integrated MXI-Express (x1), such as the NI 9154, NI 9155, NI 9157, and NI 9159, as the NI 915x.
Operate the NI 9155 only as described in these operating instructions. Safety Guidelines for Hazardous Locations The NI 9155 is suitable for use in Class I, Division 2, Groups A, B, C, D, T4 hazardous locations; Class 1, Zone 2, AEx nA IIC T4 and Ex nA IIC T4 hazardous locations;...
Page 4
PCI Express clock compatibility with the NI 9155, go to ni.com/ and enter the Info Code info 915xclock ❑ C Series I/O modules ❑ MXI-Express (x1) cable(s) up to 7 m long ❑ DIN rail mount kit (for DIN rail mounting only) ❑...
Page 5
(2.00 in.) (2.00 in.) 264.3 mm (10.41 in.) 63.5 mm (2.50 in.) 1 Measure ambient temperature here. Figure 3. NI 9155, Bottom View with Dimensions 19.0 mm Cooling Outline (0.75 in.) 50.8 mm (2.00 in.) 165.1 mm (6.50 in.) NI 9155 36.4 mm...
Page 6
Align the chassis on the surface. Fasten the chassis to the surface using two M4 or number 8 screws, as shown in Figure 6. National Instruments does not provide these screws with the chassis. NI 9155 Operating Instructions and Specifications...
Page 7
NI 9155 Figure 6. Mounting the NI 9155 Directly on a Flat Surface Make sure that no I/O modules are in the chassis before removing it from the Caution surface. Mounting the Chassis on a DIN Rail Using the NI 9915 DIN Rail Mount Kit Complete the following steps to mount the chassis on a DIN rail using the NI 9915 DIN rail mount kit.
Page 8
Figure 9 shows the mechanical dimensions of C Series I/O modules. 88.1 mm (3.47 in.) 22.9 mm 70.7 mm (0.90 in.) (2.78 in.) Figure 9. C Series I/O Module, Front and Side View with Dimensions NI 9155 Operating Instructions and Specifications ni.com...
Info Code earthground Wiring Power to the Chassis The NI 9155 requires an external power supply that meets the specifications in the Power Requirements section. The NI 9155 filters and regulates the supplied power and provides power for all of the I/O modules.
Page 11
Optionally, you can connect the positive lead of another power supply to the other V terminal and the negative lead to one of the C terminals. Install the COMBICON connector on the front panel of the NI 9155. Caution To ensure the specified EMC performance, do not connect the power input to a DC mains supply or to any supply requiring a connecting cable longer than 3 m (10 ft).
Page 12
Loads the FPGA bit stream from flash memory to the FPGA when the chassis powers on. If you want the NI 9155 to autoload and run a VI at powerup, you must also configure the VI to autoload before you compile it. For more information about autoloading VIs, refer to LabVIEW FPGA Module Help.
Link is established. Blinking yellow PCI Express clock is incompatible with NI 9155. Note For information about PCI Express clock compatibility with the NI 9155, go to and enter the Info Code ni.com/info 915xclock Powering Down the MXI-Express System Always power down the host system before powering down any connected NI 9155 chassis.
Figure 12. NI 9155 LEDs POWER LED The POWER LED is lit while the NI 9155 is powered on. This LED is a bi-color LED. When the chassis is powered from V1, the POWER LED is lit green. When the chassis is powered from V2, the POWER LED is lit yellow.
NI 9155 may require less power in your application. For more information about the I/O module power requirements, refer to the module operating instructions. Physical Characteristics If you need to clean the chassis, wipe it with a dry towel.
For additional environmental information, refer to the Minimize Our Environmental Impact web page at . This page ni.com/environment contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document. NI 9155 Operating Instructions and Specifications ni.com...
Europe (DEMKO) ........Ex nA IIC T4 Gc Where to Go for Support National Instruments corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas, 78759-3504. National Instruments also has offices located around the world to help address your support needs.
Page 20
For patents covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the NI-RIO Device Drivers Readme.