Protocol Set-Up Siemens Mammomat® 3000 - Siemens NIC2-2 Installation And Startup Manual

Network id camera
Hide thumbs Also See for NIC2-2:
Table of Contents

Advertisement

NIC2-2 · Installation and Set-up Manual
Move between the different settings by pressing up- and down arrow keys. Use the Previous
(F2) and Next (F3) keys to toggle between the alternatives.
Set the speed of the channel, 300, 600, 1200, 2400, 4800, 9600, 19200,
Speed
38400 or 57600. Factory default is 9600.
Set the parity used by the channel, NONE, EVEN or ODD. Factory default is
Parity
NONE.
Set the number of bits used by the channel, 7 or 8 bits. Factory default is 8
Databits
bits.
Here the protocol is selected. Available alternatives are:
Protocol
NONE -
MMAT3000 - Used by Siemens Mammomat® 1000/3000.
MMAT300 -
BALANCE -
PLANMED -
SENOGRAPH - Used by Senographe Models DMR, DMRV2 and Senix.
SENO800T - Used by Senographe Model T800.
I-IMAGING - Used by Instrumentarium Models Alfa, Diamond and Performa
LORAD -
GXDP-P -
GXDP-NP -
For more information, please refer to the 'Network Installation Manual'.
3.4.2.1 Protocol set-up SIEMENS Mammomat® 3000
Pressing Shift Protocol (Shift F1) when the MMAT3000 protocol is selected will bring up the
protocol set-up for the SIEMENS Mammomat® 3000 protocol..
Define text for Mammomat 3000
Focus
Anode
Filter
Grid
This window is used to define a number of text strings. The Mammomat® 3000 sends a few
parameters as 0 and 1, whereas they should be presented as texts instead. This window is
used to define those texts. For more detailed information, please refer to the Mammomat®
3000 manual.
Each string may be up to 8 characters long. Use the up and down arrows to go to the row
below or above, and use "0 text"- and "1 Text" soft-keys to select column.
Creating the picture that will be printed on the film is done in the same way as usual selecting
the COMM field type.
NOTE! The communication protocol has to be set before setting up the picture, otherwise
the COMM field will display NONE.
Example
Triacon AB Sweden
The channel is not used.
Used by Siemens Mammomat® 300.
Used by Siemens BALANCE.
Used by Planmed Models Classic, Sophie and Maxview.
Used by Lorad Model IV.
Selects the General XRAY Data Protocol with poll function.
Selects the General XRAY Data Protocol without poll function.
SF
LF
W
Mo
Rh
Mo
No grid
Grid
0 Text
1 Text
15
Publication no. 201101
Save
October 2005