3LQ
6LJQDO
1
2
3
4
5
nTELEon
6
7
nINOP
8
nDECG
9
10
nUS
11
12
nIUP
13
nTOCO
14
HR
15
TOCO
16
-
17
Gnd Analog
18
Gnd
19
nMarker
20
nFMPon
21
nFMPsig
22
-
23
IF ID
24
Gnd
25
-
Chapter 5 - Interfacing to a Fetal Monitor
Interfacing to a Fetal Monitor
The Fetal Monitor Interface on the Telemetry Receiver allows digital and
analog ouputs to the Monitor. The following table lists the signals at each
pin.
Not connected.
Not connected.
Not connected.
Not connected.
Receiver on (L).
Not connected.
DECG inop (L).
DECG Mode (L).
Not connected.
US Mode (L).
Not connected.
IUP Mode (L).
TOCO External Mode (L).
US LF or ECG Signal.
TOCO/IUP signal.
Not connected.
Analog Ground.
Tele-Marker on (L).
Tele-FMP on (L) if FMP available.
Tele-Fetal movement on (L)
Not connected
,QSXW Shorted to ground with HP 8040/41A and older Series 50 Telemetry
Interface. Open for enhanced Series 50 Telemetry.
Not connected.
Key: (L) means Logic signal is low (0 Volts).
Note— Status ouputs PIN 5, PIN 19, PIN 20 and PIN 21 have open
collector outputs, all other status signals have 5V HCMOS level
outputs.
'HWDLOV