Miele PG 8581 Supplementary Document

Miele PG 8581 Supplementary Document

Washer disinfector

Advertisement

Quick Links

Eschmann User Supplementary
Information Document
Ingenuity in
infection control
since 1830
PG 8581
Washer
Disinfector

Advertisement

Table of Contents
loading

Summary of Contents for Miele PG 8581

  • Page 1 Ingenuity in infection control since 1830 PG 8581 Washer Disinfector Eschmann User Supplementary Information Document...
  • Page 2 Information Document before the product is installed and used for the first time. This Supplementary Information Document provides sufficient information for everyday use and details the UK specific configuration for the PG 8581. This includes daily use, daily/ weekly checks and tests required for safe operation.
  • Page 3: Table Of Contents

    4 Operation 4.1 Selecting a programme 4.2 Starting a programme Appendix 1 4.3 Automatic Drying PG 8581 Dental - Components, Full Loads 4.4 Drying and Load Configurations 4.5 Programme sequence indicator Appendix 2 4.6 At the end of the programme PG 8581 Veterinary - Load Carrier and 4.7 Cancelling a program...
  • Page 4: Introduction

    The machine is designed for operation with water, salt, and additive chemicals only. Organic solvents must not be used in This PG 8581 Washer Disinfector is a Class IIb medical device the machine as under certain circumstances there is a danger of (Medical Device Regulation (EU) 2017/745).
  • Page 5: Warning And Safety Instructions

    2.4. Warning & Safety Instructions relevant factors for the process, such as chemical agents, • The standard of cleaning and disinfection in the water quality etc. are discussed with Eschmann. disinfection programmes must be routinely confirmed by the user. • Baskets and carriers should only be used for the purpose they are designed for.
  • Page 6: Overview

    Guide to the machine Overview Overview Comfort door locking mechanism Dispenser for powder cleaning agent (optional) Module slot for a communication module (Back, top right) Filter combination Test point for validation Plinth (Top, front right; only visible with lid On the back: removed) –...
  • Page 7  PC / Optical interface For opening the door before or after a programme This is used by Miele service technicians to run diagnostic checks and can also be used to update programming data in the future.  button (drying assistance)
  • Page 8: Guide To The Machine

    Guide to the machine 3.1. Guide to the machine LEDs in the buttons The buttons on the control panel have LEDs (Light Emitting Diodes) that indicate the status of the washer-disinfector. The buttons on the control panel have LEDs (Light Emitting Diodes) that indicate the status of the washer-disinfector.
  • Page 9: Using The Machine

    The time of day can optionally be displayed when the • The washer disinfector is connected to the electrical supply and is switched on (LED 9, DOOR machine is switched off. Refer to the Miele Operating Instructions. OPEN, is lit).
  • Page 10 This will have been set by the engineer at installation. Please refer to the Miele Operating Instructions.  If the emergency release is operated during a programme cycle, WARNING!
  • Page 11: Adding And Dispensing Chemicals

     Open the door to an angle of approx. 45°. This ensures that the salt flows into the reservoir more easily. 7. Clean any excess salt from the area around the salt reservoir and especially from the seal using Eschmann recommended wipes.
  • Page 12: Dos Venting

    must be filled and the dispensing system  Refilling Place the neutralising agent container neutralising agent cabinet door or on a surface which is  3.4. DOS Venting Detergent dispensed via dosing system Unscrew and remove the siphon. Plac DOS 1 - Marked blue Liquid detergent is dispensed from a 5 litre container via cabinet door.
  • Page 13 phase after the main wash (see Programme charts). The reservoir must be filled and the dispensing system vented for this to occur.  Place the neutralising agent container (red marking) on the open gent cabinet door or on a surface which is robust and easy to clean. Rinse aid dispensed via dosing system DOS 3 - Marked red ...
  • Page 14: Operation

    Operation A light touch on the relevant button is sufficient to operate the functions. The buttons can also be pressed and held for approx. 4.1. Selecting a programme 4.3. Automatic Drying 20 seconds. Display illustrations Using the shortcut buttons 1, 2 or 3 from the programme The Automatic Drying Function ensures instruments are dry All display illustrations shown in these operating instructions are list, select the Eschmann Dental or W T D programme (Wash,...
  • Page 15: At The End Of The Programme

    A light touch on the relevant button is sufficient to operate the functions. The buttons can also be pressed and held for approx. 20 seconds. 4.6. At the end of the programme Display illustrations ∧ ∨ Select Yes using the UP and DOWN buttons ( All display illustrations shown in these operating instructions are Note: The Eschmann Dental or Eschmann W T D programme examples which can be different from the actual display screens...
  • Page 16: User Maintenance

    for this purpose. User Maintenance The following need to be inspected: – All filters in the wash cabinet 5.1. Routine Checks – The spray arms in the machine and in any mobile units or baskets Before the start of each working day the user must carry –...
  • Page 17: Cleaning The Machine

     Also check that the spray arms can turn easily.  Immobile or blocked spray arms must not be used again. In this case contact Miele Service. 5.2. Cleaning the machine To remove the spray arms for cleaning: Only remove one spray arm at a time so that they are fitted into the same location.
  • Page 18: Problem Solving Guide

    Problem Solving Guide WARNING! For all fault messages: PERSONAL INJURY AND EQUIPMENT If the cause is not remedied and the machine is switched off, when it is switched back on again the fault message must be DAMAGE HAZARD acknowledged with the code. The machine will be locked for Repairs may only be carried out by Eschmann.
  • Page 19: Insufficient Salt/Water Softener

    Problem Cause and remedy Venting DOS cancelled. Venting of the dispensing system was cancelled because an insufficient flow rate has been identified. A dispensing hose may be kinked or the siphon blocked. Venting must be repeated. Check the dispensing hose for kinks or leaks. Position it so that it cannot be kinked. Check the suction aperture on the siphon for blockages and remove these as necessary.
  • Page 20: Advisory Codes (Cycle Interrupts)

    A programme has been cancelled because water intake by the machine was insufficient or severely restricted. Open the stopcocks fully Follow the further information given in the Check water inlet message in the Miele Operating Instructions. Fault 406-408 A programme was cancelled because the water intake volume is insufficient.
  • Page 21 Advisory code Cause and remedy Fault 440 The float switch in the sump of the machine has not been activated. The switch may be blocked. Remove the filter combination. Check the float switch to make sure it moves freely. The float switch is located in the sump of the machine behind the spray arm Spray arms: The programme has been cancelled because of a blocked spray arm.
  • Page 22: Process Related Faults And Messages

    6.5. Process Related Faults & Messages Problem Cause and remedy Drying during programme Drying cannot be selected at the start of a programme because drying is not available for the deactivated selected programme. Start the programme without drying. Have the drying parameters for this programme adjusted by Eschmann Wrong code entered The code entered is not the same as the code saved.
  • Page 23: Door

    6.6. Door Problem Cause and remedy Door not closed properly Protruding wash load item obstructing the door. Sort the wash load so that it does not obstruct the door. Close the door. Slamming the door can result in problems with the Comfort door lock. Open and close the door.
  • Page 24: Unsatisfactory Cleaning And Corrosion

    There are white deposits on The water softener is set too low: the wash load • Set the water softener to the correct water hardness. Refer to the Miele Operating Instructions. • There is no salt in the salt reservoir. Refill the reactivation salt.
  • Page 25: Spray Arm Monitoring/Wash Pressure

    6.8. Spray Arm Monitoring/Wash Pressure Problem Cause and remedy Spray arm monitoring - The rotation speed set has not been reached because: upper wash cabinet: Spray • Items are obstructing the machine or basket spray arms. Arrange the load so that the arm blocked or excessive spray arms can turn easily and start the programme again.
  • Page 26: Noises

    6.10. Noises Problem Cause and remedy Knocking noise in the wash • One or more spray arms are knocking against the wash load. Cancel the programme. To cabinet do this follow the instructions in Section 4.7, Cancelling a Programme. • Arrange the wash load so it cannot obstruct the spray arms.
  • Page 27: Specification

    7. Specification 7.1. Technical Specification ENVIRONMENT Maximum Relative 80% for temp up to 31°C Maximum Altitude 1500m Humidity 50% for temp up to 40°C Operating Temperature 5°C to 40°C Range MECHANICAL Height 84cm Depth 60cm Width 60cm Depth with door open 120cm Weight IWD7881 : 74kg...
  • Page 28: Components And Spares

    8793193 (Refer to Appendix 5) Wash Clean Holder (1 holder) 8793228 (Refer to Appendix 5) Washer Disinfector Log Book 119859 PG 8581 Wall Chart (A3 laminated) MUW216744 Eschmann Recommended Wipes Contact Eschmann 8.2. Spares Dr Weigert Neodisher MediClean Dental Detergent 5 Litres (Blue top)
  • Page 29: Product Disposal

    Product Disposal 9.1. Waste Electrical & Electronic Equipment (WEEE) Regulations 2006 The aim of the WEEE Regulations is to reduce the amount of waste going to landfill. All Eschmann products that must be recycled in accordance with the WEEE Regulations are marked with the “wheelie bin”...
  • Page 30: Appendix 1

    Appendix 1 PG 8581 Dental - Components, Full Loads and Load Configurations Note: The recommended programme to use for dental accessories is the Eschmann Dental cycle. HFiT 5 Cassette (95mm x 180mm, external) 8722808 White 8722804 Green 8722802 Blue 8722805...
  • Page 31 Load Configuration 1 : 346 Instruments Showing positions for: 2 x E523 carrier 18 x 8722812 HFiT 10 cassettes for 180 general instruments Detail showing cassette loaded such that water drains away Upper load carrier A102 Showing positions for: 2 x E521/1 carrier with 14 extraction forceps each 1 x E337 carrier for approx.
  • Page 32 Load Configuration 3 : 376 Instruments Showing positions for: 2 x E521/1 carrier with 14 extraction forceps each 1 x E337 carrier for approx. 140 general instruments 1 x 8732035 basket for up to 12 hinged instruments (i.e. scissors and needle holders) Upper load carrier A102 Showing positions for: 2 x E523 carrier...
  • Page 33: Appendix 2

    Appendix 2 PG 8581 Veterinary - Load Carrier and Basket Configurations Load Carriers: Load Carrier Usable Space A103 Upper carrier, top load area Width 510mm x Depth 470mm x Height 90mm. A202 Lower carrier, middle load area Width 510mm x Depth 470mm x Height 140mm.
  • Page 34 E363 - 1/6 Wire Basket Part No: M3149790 6 x Baskets can be placed in all 3 levels (18 in total). The Grille Insert for A202 must be REMOVED from the bottom loading area. E378 - 1/1 Wire Basket Part No: M3203530 1 x Basket can be placed in the bottom level only.
  • Page 35 Basket Configurations: A103 with 2 x E142s on top level A103 with 2 x 8732035s on top level A202 with 2 x E142s on middle level A202 with 2 x E142s on bottom level (Grille insert MUST be FITTED) A202 with 2 x 8732035s on middle level MUW216743-2...
  • Page 36 A202 with 2 x 8732035s on bottom level (Grille MUST be REMOVED) A202 with 1x E378 on bottom level (Grille MUST be REMOVED) Fully Loaded. A103 with 2x 8732035s on top level, 2x E142s on middle level, & 1x E378 on bottom level (Grille MUST be REMOVED) MUW216743-2...
  • Page 37: Machine Settings Table

    Appendix 3 Machine Settings Table Changes made to programmes or dispensing systems must be documented below. Initial installations settings are to be added by the Eschmann Engineer (or 'disabled' if a feature is disabled). Item Cycle Setting Confirmed By Date 'Eschmann Dental or Eschmann W T D' cycle Water Softener...
  • Page 38: Recommended Detergents & Rinse Aids

    Appendix 4 Recommended Detergents & Rinse Aids It is important to use only the recommended and validated solutions with these washer disinfectors. Each solution will work differently, and volume settings may require amending to ensure effective decontamination. - CAUTION Some chemicals not recommended by Eschmann may have a detrimental effect on the washer disinfector and/or instruments.
  • Page 39: Daily/Weekly Testing

    Appendix 5 Daily/Weekly Testing Manufacturer’s testing MUST be conducted as outlined below. We recommend the completion of our Washer Disinfector Log Book (Part number: 119858) - refer to 8.2 Spares, for ordering details. Daily testing • Clean the door seal and cabinet face using Eschmann recommended wipes (Contact Eschmann).
  • Page 40 Service Warranty Eschmann products are supported by a network of fully trained This product and purchased accessories are warranted for a engineers, offering high quality Eschmann spare parts. For further minimum period of 12 months to be free from defects in materials information on the range of Service Contracts available, please and workmanship at the time of delivery.