OPERATING INSTRUCTIONS
9. Start sterilization.
# Press the H H s f l button. A short beep sounds and "°C"lamp
lights to indicate that sterilization is in process.
The digital indicator begins to register, when the temper-
ature inside the chamber reaches 80°C.
The is
temperature inside the chamber reaches 98°C.
Note
• Sterilization can not be started unless
if it is not lit, try to close the door again.
Don't turn the handle, if
flashing.
• Turning the handle and disconnecting the wire connec-
tor, warning buzzer could sound, and §0 error Indica^
tion could be appeared on the digital display, and
lamp, IMfli lamp, liiraisMKMMia and
lamp could be flashed.
(See page 27 "Trouble Shooting Guide".)
10. Starting of sterilization timer
• When the temperature inside the chamber reaches the
sterilization temperature setting, the timer operation seg-
ments flash and the timer begins to operate.
# l f the temperature inside the chamber reaches the steri-
lization temperature setting and then drops 0.5°C(0.9°F)
or more below the setting, the timer operation segments
will light continuously and the timer will stop.
If the temperature inside the chamber reaches the steri-
lization temperature setting again the timer will start ope-
rating again.
^—( Supplementary note )-
To determine the remaining sterilization time when the
sterilization timer is operating, press the temperature set-
ting
button and the timer setting |||J5@j| button
at the same time.
23
lamp starts flashing, when the
is lit,
Door lamp
HIGH PRESSURE
lamp is
HIGH PRESSURE
(RESET)
(
START
J^ Short beep
HIGH PRESSURE
38
I MIN I
HIGH PRESSURE HI
C
I
Timer operation segments
| HIGH PRESSURE-H
|—
(Assuming a sterilization
temperature setting of 121°C)
I 3
•I L*
I MIN I
| HIGH PRESSURE-H
IU
- I MIN I -
I HIGH PRESSURE-H
|-
(10 minutes remaining)