4) Visually check the outer surface of the Magnetic sensor attachment
for leavings of the disinfectant. If necessary, repeat the rinsing.
Water, Detergent or
Disinfectant
Fig.11 Immersion of the Probe and the Magnetic sensor attachment
5.4

Drying

1) Wipe the equipment with a single-use, fluff-free wipe or towel to
remove moisture from the surface of the equipment.
2) Dry the equipment naturally in an ambient temperature between 15-30°C
for a minimum of 4 hours. Alternatively the equipment can be dried
using a drying heater at a temperature of less than 60°C.
5.5

Inspection

Inspect the equipment for any damage such as crack, scratch or
deformation. Do not use it if any damage is found.
- 21 -
Drying