Manual Cleaning:
Prepare the detergent solution in a tank with cold water (please follow
the instructions of the detergent manufacturer regarding application,
dilution and contact time).
A) C41V1 probe
1)
The temperature of the detergent solution should be between 15-30 °C,
concentration is 1.6%. Please note the minimum contact time of the
detergent in the manufacturer's instruction. If a differing
detergent is used, please also note the approved material
compatibility for the medical device.
2)
Immerse the immersible part of the probe without connector into
the diluted detergent solution (see Figure 11). Wipe the immersible
part of the probe under the surface of the detergent solution with
a soft cloth to remove all visible soil. Be sure that all grooves
of the probe are implemented during the cleaning process.
3)
The immersible part of the probe should be left in the detergent
solution according to the specified contact time of the detergent
manufacturer.
4)
Wipe the un-immersible parts of the probe with a soft cloth dipped
with the detergent solution.
5)
Rinse the probe with running tap water for 1 minute.
(alternatively: immerse the immersible part of the probe in a tray
filled with deionized water/tap water (see Fig.11) for 5 min.)
6)
Visually check the outer surface of the probe for cleanness. If
necessary, use magnifying glass for visually check. If there is
still soil visible, repeat all above steps.