- Temperature level alarm: if you choose 37.0°C (98.6°F) the thermometer alternates,
for 7 seconds, the projection of «Hi.2» and the temperature above 37.0°C (98.6°F);
if you choose 37.5°C or 38.0°C (99.5°F or 100.4°F) the thermometer alternates, for
7 seconds, the projection of «Hi.2» and the temperature above 37.5°C or 38.0°C
(99.5°F or 100.4°F).
- Air function: choose on in case of high air conditioning in the room (in case of high
air conditioning in the room, when AIR appears, wait for on and release the button -
the display will show alternately the word AIR and the temperature value); otherwise
choose oFF.
The exclusive Manual Quick Calibration System (MQCS) promptly corrects the de-
vice's internal temperature and adapts it to the real temperature of the room where the
reading is to be taken.
This system is extremely useful in case of prolonged use, when you need to keep the
temperature of the thermometer steady, or in case you have to move quickly from one
room to another – having different temperature – and you cannot wait too long to have
the thermometer stabilized to room temperature: for example, if you move from room
to room in a hospital department, especially if rooms have different exposures (North,
South, etc.) or in case of ambulance use for homecare.
This system is very useful also if the device is kept in the pocket or if handled for long
time for example during screening of passengers in airports.
Proceed as follows:
• open the protective cap and press and release
the "FACE"
and "HOME"
simultaneously: the symbol CAL will appear, and
the display will light up in blue.
• within 10 seconds focus the thermometer on
an internal wall (not the inside of an external wall)
or wardrobe with uniform temperature and at a
point approximately 80/150 cm from the floor.
Press the "HOME"
• once the right distance is reached (temperature
value between the arches, fig. 5), release the but-
ton: the lights flash slowly and the display shows
the room temperature.
To ensure a reliable temperature reading, do not
focus the thermometer on an outside wall, win-
dow, source of heating or cooling (radiator, air
conditioner, lamp, computer, surface in contact
with the human body, etc.).
8. ROOM TEMPERATURE CALIBRATION
8.1 MQCS
buttons (fig. 9)
button (fig. 10);
Fig. 9
MQCS
10
Fig. 10
MQCS