Insert the new sensor paying attention to match the sensor connectors with the relevant supports on the
6
circuit board.
Place back the tubes (See point 4).
7
Close back the base of the instrument and screw back the four screws (See point 2).
8
Turn on the instrument to check the new sensor works correctly through the menu "Sensor Troubleshooting".
It is normal if a newly installed sensor gives a 'current error': it is necessary to wait some time, so that the sensor
polarization can settle.
The table here below shows the minimum settling time for each sensor.
CODE
O
Sensor
2
Cod. AACSE50
CO Sensor with NOx filter
0-4000ppm
Cod. AACSE58
NO Sensor
Cod. AACSE60
DETECTED GAS
O
2
Oxygen
CO
Carbon Monoxide
NO
Nitric Oxide
67
POSITION
SETTLING TIME
S1
S2
S3
8 hours
2 hours
8 hours
K100000000S9 036103 130121