Figure 5-1: Detector Port Identification; Figure 5-2: Power Supply Dip Switch Settings - ABB PGC5000 Service Instruction

Gas chromatograph
Hide thumbs Also See for PGC5000:
Table of Contents

Advertisement

Figure 5-1: Detector port identification

1
2
3
4
5
Legend: Detector port identification
Item
Description
1
Purge gas inlet. (Detector assembly purge not used in most applications.)
Flameproof breather must remain open or connected to optional purge system.
2
Reaction gas inlet (plasma gas)
3
Purge gas vent. (Detector assembly purge not used in most applications.)
Flameproof breather must remain open or connected to optional purge system
4
Detector inlet (chromatograph column connection for analytes to the detector)
5
Detector vent

Figure 5-2: Power supply dip switch settings

NOTICE – Equipment damage. Dip switches on the power supply should never be
adjusted without factory support.