Setting Response Time -
AVRG
Procedure Options
2-16
Use this option to specify the number of readings that occur before
the meter will respond to a step change in flow rate. In general, the
smaller the number of readings, the less steady the display will
appear. Complete the following steps to set the response time:
1. Press [ESC], [ENTER], [ESC]. The meter enters the user program.
2. Press [ ] until PROG appears and press [ENTER].
3. Press [ ] until the desired channel appears. Press [ENTER] at CH1
or CH2.
4. Press [ ] until SETUP appears and press [ENTER].
5. Press [ ] until AVRG appears and press [ENTER].
]
6. Press [
and [ ] to select the response and press [ENTER]. For
best results, select STATS (statistics). This increases the response
time under steady flow conditions while still allowing a rapid
response time to changes in flow rate.
After completing the above steps, the user program returns to the
SET UP window. Do one of the following:
To continue regular programming, refer to Appendix A, Menu
Maps, to navigate to the desired menu.
To leave the user program, press [ESC] four times.