4.4 Jumper Selections

4.4.1 Integral Converter

For the Model 10DX3311E integral magmeter, zero return and functions associated with the
50XM1000 signal converter are established by the movable jumpers on terminal-strip J1 located in
the lower left corner of the 686B762U02 Primary Board assembly. The following table and diagram
correlates the zero return and functions with the jumper connections:
The full view of the 686B762U02 Primary Board Assembly is shown in Figure 4-3. Table 4-1 shows
selectable functions and their respective jumper settings.
JUMPER NO.
P4
P5, P6
P2
* Mutually exclusive, only one function at a time may be selected
** Indicates factory default positions
TABLE 4-1. 686B762U02 PCB JUMPER FUNCTIONS
POSITION
J1
1-2 *
3-4 *
5-6 *
7-8
9-10
1-2
2-3
J2
1-2
1-6
2-5
3-4
-
FUNCTION
Zero Return
10 KHz Signal
Reverse Pulse Signal
GND for 10 KHz, Zero Return
COMMON for Reverse Pulse
** Integral Converter
** CD-1, XM1000 Mode
** IEX = 85 to 141 mA
IEX = 125 to 230 mA
IEX = 210 to 350 mA
IEX = 337 to 560 mA
Not to be Jumpered
4-7