9.7 Patient Data

The patient data menu can be set up to meet
individual requirements. If you do not want to enter
blood pressure readings, for instance, you can
remove the corresponding prompts:
• Use the cursor keys to position the bar cursor on
"Setup Patient Data" and confirm the selection
with
.
The patient data setup menu will appear.
• Select "no" for prompts that you want to remove
from the dialog.
Items
- Last name
- First name
- Date of birth
- Patient ID
cannot be removed.
Items
- Height
- Weight
- Diastole
- Systole
- Referring Physician
- Medication
are disabled. They can be enabled from the patient
data setup menu.
• Press
to exit the menu.
48
System Setup

9.8 Option Code

In this menu you enter the option codes to enable a
number of optional software functions. The
respective option becomes active after you have
entered the code number. The code numbers are
listed on the option code sheet of the different
software options.
How to activate options
• Press
setup
The main menu appears, showing the following
items:
− Operating mode: 12 Lead (6 Lead, Arrhythmia)
− General Settings
− Communication
− Setup Patient Data
− Option Code
• Using the cursor keys, position the bar cursor on
Option Code and confirm the selection with
The option code menu appears.
• Enter the 12-digit code number from the
keyboard and exit the menu with
• Use the cursor keys to position the bar cursor on
"Option code" and confirm the selection with
The option code menu will appear.
MEAS: measurement of the 10-second resting
ECG
DIAG:
measurement and interpretation of the 10-
second resting ECG
MEMO: program for storage of up to 40 resting
ECGs
C100:
activates the three options MEAS, DIAG,
MEMO for a maximum of 100 ECGs
C500:
activates the three options MEAS, DIAG,
MEMO for a maximum of 500 ECGs
EVAL: activates the three options MEAS, DIAG,
MEMO for a maximum of 4 weeks
• Enter the code number and exit the menu with
.
MAC® 1200