Missing/Inaccurate Patient Data/Waveforms; Speaker Buzz; During Alarms; Always - GE Transport Pro Service Manual

Patient monitor
Hide thumbs Also See for Transport Pro:
Table of Contents

Advertisement

Missing/Inaccurate Patient Data/Waveforms

Speaker Buzz

During Alarms

Always

Unresponsive Buttons

Revision A
Troubleshooting: Problems
1. Refer to operator's manual for detailed information on patient
preparation and electrode placement.
2. Change the sensitivity level. Usually a lower percentage rectifies the
problem.
3. Test the patient cable on a known working device. If the patient cable
is defective, replace.
4. Test the patient connector with a patient simulator. If the patient
connector is defective, replace the acquisition device.
1. Reboot the monitor.
2. Reseat speaker cable connection.
3. Verify that speaker and/or cable is not damaged.
4. If speaker or cables are damaged, replace the speaker FRU.
5. Unplug speaker and check speaker cable resistance.
6. If resistance is between 6 and 8 ohms, replace the main processor
FRU.
7. If resistance is outside 6 and 8 ohms, replace the speaker.
1. Reboot the monitor.
2. Replace the main processor FRU.
1. Reboot the monitor.
2. Connect a known good power supply.
3. Reseat the keypad harness connection.
4. Verify that the keypad harness is not damaged.
5. If the keypad harness is damaged, replace the front bezel.
6. If the keypad harness is not damaged, replace the main processor
FRU.
Transport Pro
2012659-042
6-5