Brewer Model 5000/5001 examination during general examinations conducted by qualified table, call the Brewer Customer Service Department at medical professionals. Read and understand all safety 1.888.Brewer.1. Our trained staff will attempt to assist information, operating instructions, and maintenance you in correcting the problem directly over the phone.
SAFETY INFORMATION Safety Information The primary concern of The Brewer Company is that the Failure to follow instructions with these symbols equipment is operated and maintained with the safety of could result in serious personal injury and/or the patient and healthcare staff in mind. To ensure safe damage to the unit.
INSTALLATION Overview 1. Carefully remove outer carton by removing the staples Perform the following sequence in order when setting up along the bottom of the carton and lift off table. the table: 2. Unbolt the table from the shipping pallet by removing Unpacking the 4 bolt screws from the pallet with a 1/2”...
Page 6
INSTALLATION (CONTINUED) Leveling the Table (Figure 4) A leveling screw pad is located at each corner under the table’s base. Adjust the four leveling pads, by turning them up or down, using Method 1 or Method 2 to achieve a solid, level installation. Test step by pulling out. If step drags on floor or is difficult to pull out, adjust the front table levelers (not the step levelers) out 1 turn at a time until step pulls out easily.
OPERATION General Back Section (Figure 6) 1. Press and hold the release handle, lift back WARNING section upward to raise. Releasing the handle will automatically lock the table in the desired position. EXPLOSION HAZARD Do not use this table in an explosive or oxygen- 2.
Page 8
OPERATION (CONTINUED) Footstep (Figure 7) 1. Place hand into opening in front of step and gently lift while fully extending footstep. WARNING 2. To retract step to stowed position, gently lift while pushing step fully back into cabinet. FALL HAZARD •...
Page 9
OPERATION (CONTINUED) Stirrups (Figure 8) Retracting the stirrup to the storage position: WARNING 1. Rotate stirrup to the first lateral position (parallel to FALL HAZARD the table). Failure to engage the stirrup lateral locking 2. Fold stirrup down against bar. mechanism could allow patient to lose balance resulting in a fall and personal injury to patient.
Page 10
OPERATION (CONTINUED) Leg Extension (Figure 9) WARNING Horizontal Positioning FALL HAZARD Push leg extension in fully before allowing patient to get onto or off of table. Failure to push leg extension in fully can cause patient to fall resulting in serious injury. 1.
Page 12
OPERATION (CONTINUED) Pelvic Lift - Model 5001 Only (Figure 11) Drawer Warmer - Model 5001 Only Raising the seat section to the Pelvic Lift Position: (Figure 11) 1. Raise the front edge of the seat section. WARNING 2. Rotate the pelvic tilt support bar upward to its full BURN HAZARD upright position.
OPTIONS, ACCESSORIES, AND COMMON SERVICE PARTS Listed below are the accessories and service kits which are authorized for use with this table. Description Order No. Use/Restrictions Replacement Upholstery 5115-XX Intended Use - Replacement top with hardware for 5000 and 5001, examination tables in standard colors. Special Upholstery 5115-SP Intended Use-Allows the purchase of specialty vinyls...
Brewer warrants its Products to be free from defects in parts and workmanship under normal use and service for a period detailed in the table below from the date of delivery to original end user. Brewer will not be responsible for any Product failure due to abuse, modification or improper use, or for any use which exceeds the published capacity of the Product.