If you require assistance with the installation or operation information, and maintenance requirements contained in of your Brewer table, call the Brewer Customer Service this manual prior to operating the table. Become familiar Department at (1-888-Brewer1). Our trained staff will with all of the table functions before using it with a patient.
SAFETY INFORMATION Safety Information This is the safety alert symbol. It is used to The primary concern of The Brewer Company is alert you to potential physical injury hazards. that the equipment is operated and maintained Obey all safety messages that follow this with the safety of the patient and healthcare staff symbol to avoid possible injury or death.
Brewer FLEX Access Exam Table INSTALLATION NOTICE NOTICE Inspect Carton and Contents Proper Handling Inspect all boxes and contents for To avoid injury and damage to the table, handle only damage. Report any damage to the as indicated in Figures 2 through 16. Do not handle as carrier immediately.
Page 6
Brewer FLEX Access Exam Table INSTALLATION (continued) NOTICE In order to prepare the table for moving, you must power it up, and change the position from the position. 4. Cut the shipping strap, and discard strap and foam tube (Figure 4).
Page 7
Brewer FLEX Access Exam Table INSTALLATION (continued) 11. Press lift up on the foot control to raise the seat approximately 1” (Figure 7). 25mm 1” STOP RETURN TO CHAIR Figure 7. Raise Seat off of Brace To aid in preventing injury or product damage please observe the following cautions when removing the table from the pallet.
Page 8
Brewer FLEX Access Exam Table INSTALLATION (continued) 12. Raise the backrest to its full up position (Figure 11). STOP RETURN TO CHAIR Figure 11. Raise Backrest 13. Unplug foot control and power cord and place into front storage drawer (Figure 12.).
Page 9
Brewer FLEX Access Exam Table INSTALLATION (continued) 15. Slide table partially off of pallet and stop when there is sufficient room to place furniture dolly beneath base of table. Finish pulling table onto furniture dolly (Figure 14). Figure 14. Slide table off of pallet...
Page 10
Brewer FLEX Access Exam Table INSTALLATION (continued) Installing the Power Cord 1. Remove the power cord from the front drawer. 2. Insert the female plug end of the power cord into the receptacle on the table (Figure 17). Make Power Cord Foot Control sure the plug is fully seated.
The control system is protected by a fuse to prevent excessive current draw which could damage the system. If operation does Area Not Clear Hazard not resume after a cooling period of one hour, contact Brewer Before initiating power keep personnel and equipment clear 800-558-8777.
Page 12
Brewer FLEX Access Exam Table OPERATION (continued) GREEN LIGHT BACK UP (Lights when power (Raises the back) to the foot control is ON) RETURN TO CHAIR BUTTON STOP BUTTON (RED) (BLUE) (Stops all functions (Returns table to chair when pressed)
Page 13
Brewer FLEX Access Exam Table OPERATION (continued) Seat and Backrest Safety Switches Safety Switch Seat Switch Located on the Top Shroud. All models are equipped with switches on the seat bottom (Figure 22) which automatically stops the lowering function of the seat if any of the switches comes into contact with an object as the seat is lowering.
Page 14
Brewer FLEX Access Exam Table OPERATION (continued) Foot Pad The foot pad can be slid beneath the seat with the leg support extended for easy access to the treatment pan (Figure 25). Pushing the leg support back in will allow the glides on the foot pad to drop into slots. The foot pad will once again come out with the leg support the next time it is pulled out.
Page 15
Brewer FLEX Access Exam Table OPERATION (continued) Paper Roll Holder To change a paper roll, remove old paper roll tube, place the paper holder rod into the paper roll tube, and place into paper roll holder”J-Hooks” located on the rear of the seat back (Figure 28). Pull the end of the roll over the top of the seat back and place in the desired position.
Page 16
Brewer FLEX Access Exam Table OPERATION (continued) Adjustable Stirrups (Optional) Operating the Stirrups (Figure 31) 1. To operate the stirrups, grasp end of stirrup and pull straight out of the table. Lift up slightly while sliding the stirrup out. NOTE: Applying downward pressure when pulling stirrup from table may make it difficult to move.
Page 17
Brewer FLEX Access Exam Table OPERATION (continued) Grab Bars (Optional) In the forward (normal use) position the grab bars lock into place preventing left or right rotation. Pull up to rotate Push down to The grab bars can be rotated to the side of the table for lock in place improved access to the patient.
Page 18
Brewer FLEX Access Exam Table TABLE ADJUSTMENTS (continued) Seat Back The back can be adjusted to any angle between horizontal and 80°. CAUTION MINIMUM HEIGHT For lowering 28“ Backrest Lowering Hazard backrest to Depressing the manual backrest release horizontal position.
Page 19
Brewer FLEX Access Exam Table TABLE ADJUSTMENTS (continued) Manual Pelvic Tilt Pelvic Tilt Adjustment Bar (Models 5701 and 5801) The manual pelvic tilt feature sets the seat angle at 0° (horizontal) or 5° (up) position. Pull up on the front of the seat and the pelvic tilt automatically travels to the locked tilt position (Figure 38).
Brewer FLEX Access Exam Table MAINTENANCE Preventative Maintenance Table Care CAUTION For the latest in Cleaning Guidelines visit: www.brewercompany.com Loose Fasteners and Damaged Components Hazard Failure to perform periodic inspections of the table could result in personal injury or equipment NOTICE damage.
Page 21
Brewer FLEX Access Exam Table Control Box Fuse Replacement (for tables with the seat in the up position) WARNING WARNING Shock Hazard Disconnect power by removing plug from the power outlet before Support changing fuses. Device CRUSH HAZARD 1. Manually raise table and prop in raised position Failure to support seat using a support device intended for this purpose.
Brewer FLEX Access Exam Table OPTIONS AND ACCESSORIES Listed below are the options and accessories which are authorized for use with this table. Description Order No. Use/Restrictions Covers hole created when user chooses to remove grab Armrest Hole Cap 101872 bars.
Brewer FLEX Access Exam Table COMMON SERVICE PARTS Listed below are the common service parts and kits which are authorized for use with this table. Description Order No. Use/Restrictions Covers hole created when user chooses to remove grab Armrest Hole Cap 101872 bars.