Hitachi HF2000 TEM Operation Manual page 14

Table of Contents

Advertisement

1. Start up
Adjust the beam alignment in [ZOOM] mode.
Insert the X-ray detector.
Insert condenser aperture (2nd aperture) and minimize the swinging beam using the condenser X/Y
adjustment screws when passing through crossover.
Remove the any objective aperture and turn off the column pressure gauge filament by pressing the
"FILAMENT" switch.
Notice: not use any objective apertures when the detector is in! The objective aperture will generate
hard x-rays and burn out the detector.
2. The EDS procedure
Make sure to turn off the column pressure gauge and insert condenser aperture (2nd aperture).
Turn on the power switch of the EDS workstation.
-Screen-
Click "Spectral Display" icon.
Select "Acquire" → "Parameter"
→ "Set up". Set the max. energy to 20.48 keV.
Find region of interest and focus it properly with
brightness knob.
Notice: When the dead time is too high (over 30%),
select smaller condenser aperture, or
acquire spectrum from thinner region.
Depress
on the quick button.
Depress
, when the intensity reaches certain intensity.
Enter the file name, save the spectrum as "EMSA".
HITACHI
HF2000 TEM
Operation Manual
- EDX analysis in zoom mode -
14
EMSA形式で保存
Save as EMSA
Enter the file name
ファイル名を入力