2. Part of activities independent of serial number
Drive belt
Fig. 12 Removing the belt tightener
To tighten the drive belt, proceed as follows:
•
Remove the counterbalancing weight of the C-arm.
•
Remove the belt tightener (Fig. 12).
•
Remove the support for the drive belt (Fig. 13).
•
Increase the tension of the belt drive by one tooth.
•
Attach the support for the drive belt .
•
Tighten the drive belt .
•
Attach the counterbalancing weight for the C-arm.
•
Repeat the procedure on the other side of the C-arm.