3. Set up

3.1. Warning notices

DANGER
Escaping compressed gas
Risk of serious or even fatal injury from contact with escaping compressed gas or from unsecured plant
components.

3.1.1. Requirements for pipelines

Correctly dimensioned gaskets
Correctly aligned flanges and gaskets
Differences in pipe diameters at joints should be avoided or, if present, should not exceed 1 mm. For more information, see the
ISO 14511 standard.
Clean, properly installed pipes.

3.1.2. Requirements for the inlet/outlet pipe section

The table below shows the required inlet pipe sections with reference to the existing direction of flow.
Table of additional inlet pipe sections required
Flow obstruction upstream of measuring
section
Slight bend
(angle < 90°)
Reduction
(pipe narrows as it approaches measuring
section)
Enlargement
(pipe widens as it approaches measuring
section)
90° elbow section
or T-piece
2x 90° elbow sections
in a plane
2x 90° elbow sections
Change of direction in 3 dimensions
Shut-off valve
12 x D
15 x D
45 x D
METPOINT
SF53
®
Depressurise the system before carrying out any set up or maintenance work. Such work must be
carried out by authorised skilled technical personnel (see "Safety instructions" on Page 5).
Above 10 bar, use the high-pressure safety device for safe installation and removal.
Tighten the adapter sleeve using a torque of 20–30 N·m (spanner size 27).
Use only pressure-resistant installation materials and suitable tools that are in proper working order.
Before pressurising the system, check all device parts and repair them as necessary. Open valves
slowly to prevent pressure blow-outs during operation.
Minimum length of inlet pipe section
5 x D
5 x D
NOTE
5 x D
(L1)
12 x D
15 x D
15 x D
15 x D
20 x D
35 x D
45 x D
15 x D
5 x D
5 x D
20 x D
Deviating measurements
The above values are required minimum values (presented in a
simplified form). If the recommended settling sections cannot
be implemented, the deviations in measurements made may
become more frequent or very frequent.
Installation and operating manual
Minimum length of outlet pipe
section (L-L1)
5 x D
5 x D
5 x D
5 x D
5 x D
5 x D
5 x D
5 x D
15 x D
35 x D
5 x D
17