RENASYS TOUCH - Quick reference
(refer to the user manual for full details on RENASYS TOUCH)
1. Switching the device between continuous and intermittent therapy
- To switch between continuous and intermittent therapy, ensure therapy is paused and press the
therapy mode toggle on the home screen
Therapy mode toggle
Continuous therapy home screen
2. Starting therapy (continuous)
Step 1:
Ensure that the device is powered on
and the continuous therapy home screen is
displayed
Step 2:
use the + or – buttons to select
prescribed pressure setting
Step 3:
Press the Start/Pause therapy button
to start delivering therapy
When therapy is active the status indicator
on top of the device illuminates green and
the therapy indicator at the top of the screen
will rotate orange
3. Starting therapy intermittent
Step 1:
Ensure that the device is powered
on and the intermittent therapy home screen
is displayed
Step 2:
Change the High and Low pressure
settings and cycle times by selecting the
setting and use the + or – to the prescribed
pressure/time
Step 3:
Press the Start/Pause therapy
button to start delivering therapy
When therapy is active the status indicator
on top of the device illuminates green and
the therapy indicator at the top of the screen
will rotate orange
Intermittent therapy home screen
Therapy set point
Decrease value
Therapy mode toggle
All 4 settings can be adjusted within the following pressure
ranges:
High Therapy
25, 40, 50, 60, 70, 80, 90, 100, 120, 140, 160, 180, 200mmHg
High Cycle times:
3, 5, 8, 10 minutes
Low Therapy
0, 25, 40, 50, 60, 70, 80, 90, 100, 120, 140, 160, 180mmHg
Low Cycle times:
2, 3, 5, 8, 10 minutes
Increase value
Status indicator
Start/Pause therapy
Status indicator