GE BiliSoft 2.0 Operation Manual page 20

Phototherapy system
Table of Contents

Advertisement

Component/Control
1
1
1-2
1. Standby Switch – Turns the
unit ON/OFF. The green light in
the switch indicates that the
standby switch is turned on
and the unit has power.
2
2. Hour Meter –The non-
resettable hour meter shows
3
the number of hours a light pad
has been active during the life
4
of the light box. If the Light Pad
connector is not fully inserted
5
into the light box, the LED is
automatically shut off and the
hour meter does not run. Note:
6
The hour meter is provided to
track LED life and is not
intended to be used to measure
therapy durations.
3. Unit Overheated Indicator –
When this red indicator light is
ON, the unit has overheated.
See Troubleshooting Guide for
more details.
2
4. LED Module Failure Indicator -
When the red indicator light
FLASHES, the LED module has
3
failed. See the Troubleshooting
Guide for more details.
4
5. Light Pad Connector Port –
5
Where the light pad assembly
connects to the main unit. A
6
shutoff switch inside the port
automatically turns off the LED
module whenever the fiberoptic
light pad assembly is removed.
6. Exhaust Air Vents – For proper
cooling it is important to keep
the air vents clear of
obstructions.
Function