Note:
Shut down the system
Follow the steps below to shut down the system.
Step
1
2
ÄKTA go Operating Instructions 29360951 AC
If you have a column valve, make sure all positions are put in storage solution.
Action
Select Exit UNICORN from the File menu in any module in the UNICORN
software.
Press the On/Off button on the instrument control panel. The circular illumi-
nation flashes with white color while the instrument is shutting down.
WARNING
Power is still supplied to some internal electronics circuits when
the instrument is switched off using the On/Off button. Disconnect
the instrument from the power supply before maintenance or ser-
vice.